GORE VIABIL Biliary Endoprosthesis VSWVH1010Recalls & reported problems
W. L. Gore & Associates, Inc. · VIABIL SHORT WIRE BILIARY ENDO 10mmX10cmX200cm 8.5Fr · Class 2 · Implant
All models of GORE VIABIL Biliary Endoprosthesis (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132638819), received between Aug 1, 2019 and Oct 1, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K170740); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Device Dislodged or Dislocated1
- Off-Label Use1
Patient problems
- Unspecified Hepatic or Biliary Problem2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 1, 2024Fracture; Fracture
IT WAS REPORTED TO GORE BY CONMED (CEN-74960), DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE USING THE GORE® VIABIL® SHORT WIRE BILIARY ENDOPROSTHESIS, DURING DEPLOYMENT OF A 10X10 VIABIL, THE STENT BEGAN TO "CANDY CANE" INSIDE OF THE BILE DUCT. THE "CANDY CANE" WAS FULLY DEPLOYED AND BROKE INTO THREE PIECES. THERE WAS A 2-HOUR DELAY IN THE PROCEDURE TO RETRIEVE THE PIECES. THERE WAS NO INDICATION OF PATIENT/USER IMPACT.
Patient problems: Unspecified Hepatic or Biliary Problem
MAUDE report 20347680 ↗MalfunctionJul 25, 2022Fracture; Off-Label Use; Fracture; Off-Label Use
IT WAS REPORTED TO GORE BY (B)(6) THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE USING THE GORE® VIABIL® BILIARY ENDOPROSTHESIS, THE SURGEON WAS ATTEMPTING TO REMOVE A VIABIL STENT THAT WAS PLACED ON (B)(6) 2022. REPORTEDLY, THE STENT FELL APART AND ONLY HALF OF IT CAME OUT WHEN BEING REMOVED. IT WAS REPORTED THE SURGEON DECIDED TO KEEP HALF THE STENT (DISTAL END) WITHIN COMMON BILE DUCT DUE TO THE BILE STILL BEING ABLE TO FLOW FREELY. REPORTEDLY, THIS STENT INCLUDED THE CYSTIC DUCT DRAINAGE HOLES WHICH WAS LIKELY THE CAUSE OF EXCESS TISSUE INGROWTH WHICH RESULTED IN THE STENT BREAKING APART DURING REMOVAL. IT WAS REPORTED THE PATIENT WAS NOT HARMED AND NO SURGICAL INTERVENTION WAS NEEDED. REPORTEDLY, THERE ARE NO IMAGES AVAILABLE AND THE ANATOMICAL LOCATION OF THE DEVICE WAS THE COMMON BILE DUCT. IT WAS REPORTED A STANDARD CONMED FX WIRE .035 GUIDEWIRE AND DUODENOSCOPE WAS USED. REPORTEDLY, THIS STENT WAS ALREADY IN THE PATIENT TO TREAT A STRICTURE AND WAS BEING REMOVED. IT WAS REPORTED THE PHYSICIAN DID NOT RECALL ANY OUT THE ORDINARY MANIPULATION OF THE ENDOSCOPE OR CATHETER DURING THE CASE. REPORTEDLY, THE PHYSICIAN WAITED A FEW MINUTES AFTER ATT
Patient problems: Unspecified Hepatic or Biliary Problem
MAUDE report 15094555 ↗InjuryAug 1, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED TO GORE BY CONMED (CEN 5111) THAT A GORE® VIABIL® BILIARY ENDOPROSTHESIS WAS IMPLANTED IN A PATIENT DURING AN ERCP PROCEDURE. IT WAS REPORTED THE DEVICE MIGRATED EIGHT WEEKS LATER. IT WAS REPORTED THE STENT MOVED UP INTO THE DUCT AND WAS NO LONGER TRANSPAPILLARY. ADDITIONALLY, IT WAS REPORTED THE PATIENT WAS SEEN ON (B)(6) 2019 FOR STENT MANAGEMENT WHEN IT WAS OBSERVED THE STENT HAD MIGRATED APPROXIMATELY 5-7 MILLIMETERS PROXIMALLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8853609 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170740. One submission often covers many models, so check the product and lot details in each record.
- Z-0158-2027 · Sep 3, 2026 · Open, ClassifiedGeneric Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006 Product Description: The GORE VIABIL Short Wire Biliary Endoprosthesis
- Z-2915-2026 · May 22, 2026 · Open, ClassifiedGORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may b
- Z-2914-2026 · May 22, 2026 · Open, ClassifiedGORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may b
- Z-1179-2024 · Jan 19, 2024 · Open, ClassifiedGORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes
Other Catheter, Biliary, Diagnostic devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.