Skip to content

GORE VIABIL Biliary Endoprosthesis VSWVN1006Recalls & reported problems

W. L. Gore & Associates, Inc. · VIABIL SHORT WIRE BILIARY ENDO 10mmX6cmX200cm 8.5Fr · Class 2 · Implant

All models of GORE VIABIL Biliary Endoprosthesis (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132638864), received between Jan 30, 2025 and Jan 30, 2025 — classified as 1 malfunction.

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K170740); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 3, 2026Open, Classified

    The potential for the devices to be mislabeled with the incorrect length.

    Z-0158-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2025

    STENT WAS INSERTED INTO SCOPE AND WOULD NOT DEPLOY. THERE IS A SPACE AT THE TIP OF DEPLOYER AND SHEATH WHICH MAY BE THE CAUSE OF THE DEFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272682 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170740. One submission often covers many models, so check the product and lot details in each record.

  • Z-2915-2026 · May 22, 2026 · Open, Classified
    GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may b
  • Z-2914-2026 · May 22, 2026 · Open, Classified
    GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may b
  • Z-1179-2024 · Jan 19, 2024 · Open, Classified
    GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes

Other Catheter, Biliary, Diagnostic devices

All Catheter, Biliary, Diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page