Skip to content

DeRoyal M2200-CP-CS2Recalls & reported problems

Deroyal Industries, Inc. · Chair Sensor Pad · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00749756980456), received between Dec 10, 2021 and Dec 10, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “No Audible Alarm” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2021No Audible Alarm; No Audible Alarm

    CHAIR SENSOR PAD FOR CHAIR ALARM FOR FALL PRECAUTIONS ARE NOTED TO FAIL WHEN THE PATIENT STANDS AND THE ALARM DOES NOT GO OFF. CHAIR SENSOR PAD AND CONNECTIVITY POINT TO THE PORTABLE CHAIR ALARM BOX APPROPRIATELY ENGAGED, BATTERIES ARE OK. TESTING WITH A NEW CHAIR PAD TO THE CHAIR ALARM BOX WORKS WELL. ADDITIONALLY, IF CHAIR SENSOR PAD GETS WET FROM INCONTINENCE THROUGH INCONTINENCE PADS ON THE CHAIR (OVER THE CHAIR SENSOR PAD), THE CHAIR SENSOR WILL ALSO NOT WORK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12976455 ↗

Other MONITOR, BED PATIENT devices

All MONITOR, BED PATIENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page