Skip to content

NA HY6Y97R26Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK HY6Y97R26 HCAGRAND · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 12, 2022Class IIOpen, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1028-2023 · Root cause: Process change control · Recall details · FDA record ↗

Other Pump, blood, cardiopulmonary bypass, non-roller type devices

All Pump, blood, cardiopulmonary bypass, non-roller type devices →