Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00763000059620), received between May 10, 2018 and Oct 7, 2024 — classified as 1 death, 7 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Pacing Problem” (3 reports).

16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Appropriate Device Problem Term/Code Not Available3
  • Pacing Problem3
  • Battery Problem2
  • Device Operates Differently Than Expected1
  • Electrical /Electronic Property Problem1
  • Positioning Problem1
  • Reset Problem1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Unspecified Infection4
  • Atrial Flutter1
  • Bradycardia1
  • Death1
  • Erosion1
  • Muscle Weakness/Atrophy1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2024Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED "HEART RATE IN THE 40S" AND "VOMITING". THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Vomiting

    MAUDE report 20382951 ↗
  • InjuryJul 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS REMOVED AND REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 19679009 ↗
  • InjuryJan 19, 2024Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED WEAKNESS AND TWITCHING. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) 'FIRED'. UPON INTERROGATION, IT WAS NOTED THAT THE IPG WAS WORKING APPROPRIATELY. THE PATIENT WAS HOSPITALIZED AND THEIR HOSPITALIZATION PERIOD WAS EXTENDED. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Muscle Weakness/Atrophy; Twitching

    MAUDE report 18542143 ↗
  • MalfunctionJul 14, 2022Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) CARDIAC COMPASS HAD INVALID DATA. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15011439 ↗
  • InjuryMar 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS IPG. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 13665575 ↗
  • InjuryJul 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN EROSION. IT WAS NOTED THAT A SMALL SECTION ON THE IMPLANTABLE PULSE GENERATOR (IPG) HAD ERODED THROUGH THE SKIN. THE IPG WAS EXPLANTED AND REPLACED WITH A NEW IPG IMPLANTED IN A SUBMUSCULAR LOCATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Erosion

    MAUDE report 12171844 ↗
  • MalfunctionJun 2, 2021Reset Problem; Reset Problem

    IT WAS FURTHER REPORTED THAT THE DEVICE WAS REPROGRAMMED.

    Patient problems: Atrial Flutter

    MAUDE report 11918916 ↗
  • MalfunctionFeb 10, 2020Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A LARGE DROP IN LONGEVITY ESTIMATION. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9686869 ↗
  • MalfunctionDec 10, 2019Pacing Problem; Pacing Problem

    IT WAS REPORTED BY A HEALTHCARE PRACTITIONER THAT THE IMPLANTABLE PULSE GENERATOR (IPG) IS MEASURING THE PATIENTS HEART RATE IN THE FORTIES. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia

    MAUDE report 9443048 ↗
  • InjuryJun 10, 2019Pacing Problem; Pacing Problem

    IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED SYNCOPE. THE PATIENT ALSO REPORTED THAT THEIR HEART RATE WAS LOW, "DOWN TO THE HIGH 30'S / 40'S ". THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Syncope

    MAUDE report 8682188 ↗
  • InjuryMay 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) AND RIGHT VENTRICULAR (RV) LEAD WERE EXPLANTED DUE TO INFECTION. A NEW IPG AND RV LEAD WERE IMPLANTED AT A LATER DATE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 8630918 ↗
  • MalfunctionApr 23, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE PATIENT THAT THEY HAVE A PROBLEM WITH THEIR IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8538937 ↗
  • MalfunctionOct 31, 2018Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) BATTERY ESTIMATE WAS DECREASING FASTER THAN EXPECTED. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8027002 ↗
  • DeathSep 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE FAMILY MEMBER THAT THE PATIENT DEVELOPED AN INFECTION APPROXIMATELY THREE WEEKS POST IMPLANTABLE PULSE GENERATOR (IPG) AND RIGHT VENTRICULAR (RV) LEAD IMPLANT PROCEDURE AND DIED. ADDITIONAL INFORMATION OBTAINED REPORTED THE CAUSE OF DEATH AS POCKET INFECTION WITH ASSOCIATED SEPSIS.

    Patient problems: Death; Sepsis; Unspecified Infection

    MAUDE report 7896526 ↗
  • InjuryAug 7, 2018Positioning Problem; Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING USE OF THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM, THE PATIENT COMPLAINED OF PAIN AT THE SITE OF THE PACEMAKER. X-RAY REVEALED THAT THE LEAD SLACK THAT WAS PRESENT POST IMPLANT WAS WRAPPED AROUND THE IPG IN THE POCKET AND THE PACEMAKER POSITIONING HAD CHANGED, WITH APPEARANCE TO HAVE BEEN MANIPULATED IN THE POCKET. UPON REQUEST, THE IPG SYSTEM WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain; Twiddlers Syndrome

    MAUDE report 7759685 ↗
  • MalfunctionMay 10, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    THE PATIENT REPORTED THAT THEY WERE UNSURE IF THEIR IMPLANTABLE PULSE GENERATOR (IPG) WAS WORKING PROPERLY OR NOT. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7503062 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-1924-2025 · Apr 29, 2025 · Open, Classified
    Azure S DR MRI SureScan, Product number W3DR01
  • Z-2508-2020 · Jun 4, 2020 · Open, Classified
    Azure S DR MRI, Model Number W3DR01
  • Z-2509-2020 · Jun 4, 2020 · Open, Classified
    Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
  • Z-2510-2020 · Jun 4, 2020 · Open, Classified
    Azure S SR MRI, Model Number W3SR01
  • Z-2507-2020 · Jun 4, 2020 · Open, Classified
    Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
  • Z-0806-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (
  • Z-0809-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
  • Z-0810-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
  • Z-0807-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Numb
  • Z-0811-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
  • Z-0808-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
  • Z-2120-2011 · Apr 4, 2011 · Terminated
    Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
  • Z-1440-2010 · Feb 11, 2010 · Terminated
    Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,

…and 1 more.

Other Implantable pulse generator, pacemaker (non-CRT) devices

All Implantable pulse generator, pacemaker (non-CRT) devices →

Sources and updates