Fortrex™ A35HPV06100080 (UDI 00763000106041)Recalls & reported problems

Covidien LP · PTA A35HPV06100080 FORTREX V02 · Class 2

All models of Fortrex (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000106041), received between Aug 11, 2021 and Jan 3, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Burst Container or Vessel1
  • Inflation Problem1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Thrombosis/Thrombus2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING INDEX PROCEDURE A ELLIPSYS CATHETER WAS USED TO TREAT A LESION LOCATED IN THE LEFT ARM AND TWO FORTREX BALLOONS DURING THE SECONDARY PROCEDURE. APPROXIMATELY 5 MONTHS POST INDEX PROCEDURE PATIENT SUFFERED A CLOTTED AV FISTULA. THE PATIENT WAS TREATED WITH PERCUTANEOUS INTERVENTION. PATIENT RECOVERED.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 18433841 ↗
  • InjuryJan 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING INDEX PROCEDURE A ELLIPSYS CATHETER WAS USED TO TREAT A LESION LOCATED IN THE LEFT ARM AND TWO FORTREX BALLOONS DURING THE SECONDARY PROCEDURE. APPROXIMATELY 5 MONTHS POST INDEX PROCEDURE PATIENT SUFFERED A CLOTTED AV FISTULA. THE PATIENT WAS TREATED WITH PERCUTANEOUS INTERVENTION. PATIENT RECOVERED.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 18433837 ↗
  • MalfunctionDec 14, 2021Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    PHYSICIAN ATTEMPTED TO USE AN FORTREX HP PTA BALLOON CATHETER DURING PATIENT TREATMENT. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. AN INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT. IT IS REPORTED BALLOON INFLATION DIFFICULTIES OCCURRED WHEN DEVICE WAS INFLATED TO NOMINAL BURST PRESSURE. THERE WAS NO LEAK NOTED IN THE DEVICE. THE BALLOON DID NOT BURST. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. A NEW FORTREX DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13003890 ↗
  • MalfunctionAug 11, 2021Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN WAS ATTEMPTING TO USE A FORTREX PTA BALLOON DURING PROCEDURE TO TREAT A MODERATELY CALCIFIED LESION IN THE PROXIMAL CENTRAL VENOUS FISTULA WITH 70% STENOSIS. THE VESSEL WAS MODERATELY TORTUOUS. THERE WAS NO DAMAGE NOTED TO THE PACKAGING AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED AS PER IFU WITH NO ISSUES. THERE WAS NO RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE. IT WAS REPORTED THAT THE BALLOON BURST LONGITUDINALLY DURING ONE INFLATION AT 10 ATM. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER BALLOON. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12306704 ↗