Skip to content

Fortrex™ A35HPV08040135Recalls & reported problems

Covidien LP · PTA A35HPV08040135 FORTREX V02 · Class 2

All models of Fortrex (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000106201), received between May 27, 2024 and May 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2024Nonstandard Device; Device Markings/Labelling Problem; Nonstandard Device; Device Markings/Labelling Problem

    A PHYSICIAN USED A FORTREX HP PTA BALLOON WITH A 6FR SHEATH AND AN 035 NON-MEDTRONIC GUIDEWIRE DURING TREATMENT OF THE PATIENTS BILIARY ARTERY. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. EMBOLIC PROTECTION WAS NOT USED. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITH NO ISSUES NOTED. A NON-MEDTRONIC INFLATION DEVICE WAS USED, CONNECTED AS PER IFU WITH A 50% CONTRAST 50% SALINE USED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED AN 8MM BALLOON WAS UTILIZED, BUT THERE SEEMED TO BE A DISCREPANCY BETWEEN THE SIZE INDICATED AND THE ACTUAL SIZE PERCEIVED (6MM). PHYSICIAN INFLATED BALLOON UP TO 20ATM BUT STILL DID NOT REACH THE SIZE OF 8MM. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. PATIENT ALREADY DISCHARGED. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19401008 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates