NanoCross™ Elite AB14W020060150Recalls & reported problems
Ev3, Inc · PTA AB14W020060150 NANOCROSS 014 V05 · Class 2
All models of NanoCross Elite (63)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000109462), received between Oct 30, 2024 and Oct 30, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A NANOCROSS PTA BALLOON WAS BEING USED FOR TREATMENT OF A 60MM CALCIFIED LESION WITH 100% STENOSIS IN THE MID RIGHT SUPERFICIAL FEMORAL ARTERY. THE ARTERY WAS 6MM IN DIAMETER WITH SEVERE CALCIFICATION AND MINIMAL TORTUOSITY. CHRONIC TOTAL OCCLUSION SUB-INTIMAL. A 7FR NON MEDTRONIC SHEATH WITH A 0.014" NON MEDTRONIC GUIDEWIRE WERE USED. THE IFU WAS FOLLOWED. THE VESSEL WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. SEVERE RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT OF THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT IS NOT KNOWN HOW THE BALLOON WAS INFLATED, 50/50 CONTRAST INFLATION FLUID WAS USED. IT WAS REPORTED AT THE TARGET LESION THE BALLOON TIP BROKE OFF IN THE VESSEL, THE DETACHED COMPONENT WAS NOT REMOVED IT WAS TACKED TO THE VESSEL WALL. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS ABORTED.
Patient problems: Insufficient Information
MAUDE report 20567934 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.
- Z-0904-2025 · Dec 10, 2024 · Open, ClassifiedMedtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page