NanoCross™ Elite AB14W030120150 (UDI 00763000109844)Recalls & reported problems
Ev3, Inc · PTA AB14W030120150 NANOCROSS 014 V05 · Class 2
All models of NanoCross Elite (63)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000109844), received between Apr 5, 2024 and Jun 27, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem1
- Inflation Problem1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2025Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING TREATMENT IN THE PATIENTS RIGHT MID TIBIAL/POPLITEAL TRUNK USING THE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY AB14W030120150 NANOCROSS 014 BALLOON, THERE WERE DEFLATION DIFFICULTIES. THERE WAS NO DAMAGE NOTED TO THE BALLOON CATHETER. THERE WERE NO ISSUES NOTED DURING INFLATION. THE BALLOON TOOK 5 MINUTES TO FULLY DEFLATE. THE PROCEDURE INVOLVED USE OF A 6FR X45 NON-MEDTRONIC SHEATH AND A .014 NON-MEDTRONIC GUIDEWIRE, WITH INFLATION PERFORMED USING A NON-MEDTRONIC INFLATION DEVICE AND 50/50 CONTRAST FLUID. THE BALLOON WAS DEFLATED USING NEGATIVE PRESSURE WITH A 60 CC SYRINGE. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22344036 ↗MalfunctionApr 5, 2024Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
PHYSICIAN ATTEMPTED TO USE A NANOCROSS ELITE PTA BALLOON WITH A 6X45 NON-MEDTRONIC SHEATH AND A NON-MEDTRONIC GUIDEWIRE DURING TREATMENT OF A 120MM CALCIFIED LESION IN THE PATIENTS LEFT MID ANTERIOR TIBIAL ARTERY. MINIMAL VESSEL TORTUOSITY AND MODERATE VESSEL CALCIFICATION WERE REPORTED. LESION EXHIBITED 30% STENOSIS. ARTERY DIAMETER REPORTED AS 3MM. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU (INSTRUCTION FOR USE) WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE WITHOUT ISSUE. EMBOLIC PROTECTION WAS NOT USED. THE VESSEL WAS NOT PRE-DILATED. THE VESSEL WAS POST-DILATED. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. 50/50 CONTRAST INFLATION FLUID WAS USED. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT OF THE DEVICE. INFLATION DIFFICULTIES DURING PROCEDURE WERE REPORTED. A LEAK WAS NOTED ON THE SHAFT OF THE DEVICE. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. A REPLACEMENT NON-MEDTRONIC DEVICE WAS USED TO COMPLETE PROCEDURE. NO PATIENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19049187 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.
- Z-0904-2025 · Dec 10, 2024 · Open, ClassifiedMedtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page