Skip to content

NanoCross™ Elite AB14W035040150Recalls & reported problems

Ev3, Inc · PTA AB14W035040150 NANOCROSS 014 V05 · Class 2

All models of NanoCross Elite (63)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000109929), received between Oct 7, 2021 and Oct 27, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2023Burst Container or Vessel; Burst Container or Vessel

    A PHYSICIAN WAS ATTEMPTING TO USE A NANOCROSS PTA BALLOON FOR TREATMENT OF A 150MM CALCIFIED LESION WITH CHRONIC TOTAL OCCLUSION IN THE MID LEFT TIBIAL/POPLITEAL TRUNK <(>&<)> POSTERIOR TIBIAL ARTERY. THE ARTERY WAS 3.5MM IN DIAMETER WITH MODERATE TORTUOSITY AND SEVERE CALCIFICATION. A 6FR NON MEDTRONIC SHEATH WAS USED. A SPIDER WAS USED FOR EMBOLIC PROTECTION. A NON MEDTRONIC INFLATION DEVICE WAS USED. THE DEVICE WAS PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. IT WAS REPORTED PREDILITATED WITH 2.5MM X150 NANO CROSS. ONCE 3.5X40 NANO CROSS WAS INFLATED A PINHOLE BURST OCCURRED AT 14 ATM AT THE RE-ENTRY POINT WHERE IT WAS SEVERELY CALCIFIED. PTA WAS RETRIEVED AND NO DAMAGE WAS DONE TO THE ARTERY. THE DEVICE HAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS FELT DURING ADVANCEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18019744 ↗
  • MalfunctionOct 7, 2021Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN WAS ATTEMPTING TO USE THE NANOCROSS PTA BALLOON DURING PROCEDURE. THERE WAS NO DAMAGE NOTED TO PACKAGING AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE BALLOON WAS INFLATED WITH AN INFLATION DEVICE. THE DEVICE WAS PREPPED PER THE IFU WITH NO ISSUES. THERE WERE INFLATION DIFFICULTIES. IT WAS REPORTED THAT THE BALLOON BURST DURING BALLOON INFLATION AT 4ATM. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER BALLOON. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12589950 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.

  • Z-0904-2025 · Dec 10, 2024 · Open, Classified
    Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page