NanoCross™ Elite AB14W050100150Recalls & reported problems
Ev3, Inc · PTA AB14W050100150 NANOCROSS 014 V05 · Class 2
All models of NanoCross Elite (63)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000110307), received between Apr 24, 2019 and Mar 14, 2023 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel3
- Leak/Splash2
- Deflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Patient Problem/Medical Problem1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 14, 2023Burst Container or Vessel; Leak/Splash; Burst Container or Vessel; Leak/Splash
A PHYSICIAN ATTEMPTED TO USE TWO NANOCROSS BALLOONS DURING TREATMENT OF CALCIFIED LESIONS WITH CHRONIC TOTAL OCCLUSION IN THE SUPERFICIAL FEMORAL ARTERY (SFA) AND THE TIBIAL/POPLITEAL TRUNK (TPT). THE DIAMETER OF THE SFA WAS 5MM AND THE TPT WAS 3MM. THE LENGTH OF THE LESION WAS 20-40MM. THE DEGREE OF TORTUOSITY AND CALCIFICATION WAS MODERATE. A 6FR NON MEDTRONIC SHEATH AND A NITRIX GUIDEWIRE WERE USED. THE DEVICES WERE PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. THE BALLOONS WERE INFLATED USING A NON MEDTRONIC DEVICE WITH HEP SALINE 60/CONTRAST 40. IT WAS REPORTED THE PHYSICIAN INITIALLY USED A 3MM BALLOON TO PRE DILATE THE CHRONIC TOTAL OCCLUSION IN THE SFA. HAWKPLUS M WAS USED SUCCESSFULLY. PHYSICIAN ATTEMPTED TO INFLATE A 5MM NANOCROSS BALLOON AND IT RUPTURED AT 8 ATM DURING THE INITIAL INFLATION AND WAS REMOVED INTACT. THEN ATTEMPTED TO INFLATE A 3MM BALLOON IN THE TPT UNSUCCESSFULLY, IT DID NOT REGISTER PRESSURE AND HAD ONE INFLATION AT 8 ATM FOR TWO MINUTES AND WAS REMOVED INTACT. THEY THEN USED 2.5-100 NANOCROSS SUCCESSFULLY IN TPT AND DEPLOYED AN ENTRUST STENT. THE SFA WAS POST DILATED WITH A NANOCROSS AT 8 ATM FOR 2 MINS. NO PATIENT INJURY WAS REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16535931 ↗MalfunctionNov 24, 2022Burst Container or Vessel; Leak/Splash; Burst Container or Vessel; Leak/Splash
A NANOCROSS ELITE PTA BALLOON CATHETER WAS BEING USED FOR THE TREATMENT OF AN UNKNOWN LESION IN THE POPLITEAL ARTERY. THE BALLOON WAS INFLATED USING THE EVEREST INFLATION DEVICE. IT WAS REPORTED THE BALLOON RUPTURED/LEAKED AT 8ATM. THE PROCEDURE ENDED AFTER THE EVENT. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15859968 ↗MalfunctionMay 10, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE A NANOCROSS ELITE PTA BALLOON DEVICE TO TREAT A RIGHT MID POPLITEAL ARTERYLESION IN A PATIENT THAT¿S FIBROUS WITH 70% STENOSIS AND LITTLE CALCIFICATION. A 0.014 NON MEDTRONIC WIRE WAS USED WITH A NON-MEDTRONIC SHEATH AND NON-MEDTRONIC INFLATION DEVICE. 50/50 INFLATION FLUID WAS USED. THERE WAS NO DAMAGE NOTED TO PACKAGING (SHELF CARTON, HOOP/TRAY). THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED AS PER IFU. IT WAS REPORTED THAT THE BALLOON BURST OCCURRED DURING INFLATION AT 4ATMS, THE BALLOON WAS INFLATED 4 TIMES IN THE SAME AREA WITH NO ISSUES, RUPTURE OCCURRED AT 5TH INFLATION. DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. THE PROCEDURE WAS COMPLETED WITH AA NEW BALLOON. NO PATIENT INJURY REPORTED FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11798354 ↗InjuryApr 24, 2019Deflation Problem; Deflation Problem
THE PHYSICIAN WAS INTENDING TO USE A NANOCROSS ELITE TO TREAT A 100MM CALCIFIED PLAQUE LESION WITH 90% STENOSIS IN THE LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). ARTERY DIAMETER OF 5MM. A 7FR NON-MEDTRONIC SHEATH AND A 5MM SPIDERFX WAS USED IN THE PROCEDURE. THERE WAS NO DAMAGE TO THE DEVICE PACKAGING AND NO ISSUES WHEN REMOVING THE DEVICE FROM ITS PACKAGING. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT AND EXCESSIVE FORCE WAS NOT USED. THE BALLOON WAS INITIALLY INFLATED USING A NON MEDTRONIC INFLATION DEVICE WITH ISOVUE 370 HEPARINIZED SALINE FLUID. IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES WERE ENCOUNTERED AND THE DEVICE WOULD NOT DEFLATE AT THE LESION SITE. THE BALLOON WAS DEFLATED USING A WIRE PUNCTURE TECHNIQUE THROUGH THE BALLOON LUMEN. THE DEFLATED BALLOON WAS REMOVED FROM THE PATIENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 8543480 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.
- Z-0904-2025 · Dec 10, 2024 · Open, ClassifiedMedtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page