Skip to content

RapidCross™ A14BX020150170 (UDI 00763000110697)Recalls & reported problems

Covidien LP · PTA A14BX020150170 RAPIDCROSS V04 · Class 2

All models of RapidCross (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000110697), received between Apr 14, 2022 and Sep 11, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Burst Container or Vessel1
  • Deflation Problem1
  • Difficult to Remove1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A RAPIDCROSS PTA BALLOON WAS BEING USED FOR TREATMENT. THE IFU WAS FOLLOWED. IT WAS REPORTED THE BALLOON DETACHED AT THE TARGET LESION. THE DETACHED BALLOON WAS REMOVED WITH A SNARE, THE DEVICE WAS SAFELY REMOVED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20191053 ↗
  • InjurySep 24, 2022Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component

    IT WAS REPORTED THAT A RAPIDCROSS PTA 2.0MM X 150MM X 170CM  BALLOON WAS USED TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION WITH 100% STENOSIS IN THE MID POSTERIOR TIBIAL ARTERY. IT WAS REPORTED THAT  A 6 FR COOK NON MEDTRONIC SHEATH AND A 014 CHOICE NON MEDTRONIC GUIDEWIRE WAS USED. EMBOLIC PROTECTION WAS NOT USED. THE BALLOON WAS INFLATED WITH A NON MEDTRONIC INFLATION DEVICE. HALF/HALF CONTRAST /SALINE INFLATION FLUID WAS USED. THE DEVICE WAS PREPPED PER IFU. THE DEVICE PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. THE BALLOON BURST AT 8 ATMS ON THE FIRST INFLATION. DISTAL PORTION OF BALLOON REMAINED IN VESSEL UPON RETRIEVING AND GOOSE NECK SNARE WAS UTILIZED TO ATTEMPT RETRIEVAL. DETACHMENT OCCURRED UPON REMOVAL ATTEMPT. THE BREAK/FRACTURE OCCURRED AT THE BALLOON. THE DETACHED PORTION OF THE DEVICE REMAINED IN THE PATIENT. THE PROCEDURE WAS COMPLETED BY PLACING A 4.0 COOK ZILVER STENT IN PERONEAL UP TO TPT TO PREVENT PROXIMAL PORTION OF BALLOON FROM EMBOLIZING/MIGRATING. NO FURTHER PATIENT INJURY.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient

    MAUDE report 15475825 ↗
  • MalfunctionApr 14, 2022Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    PHYSICIAN WAS ATTEMPTING TO USE A RAPIDCROSS PTA BALLOON DURING PROCEDURE TO TREAT A MODERATELY CALCIFIED PLAQUE LESION IN THE MID A NTERIOR TIBIAL ARTERY (ATA). THE VESSEL WAS LITTLE TORTUOUS. NO EMBOLIC PROTECTION WAS USED. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUE NOTED WHEN REMOVING DEVICE FROM HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. BALLOON DEFLATION DIFFICULTIES WAS REPORTED WITH DEVICE NOT ABLE TO DEFLATE AT LESION SITE. THE DEFLATION ISSUE WAS NOTED FOLLOWING THE FIRST INFLATION. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THERE WAS NO RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE. IT WAS REPORTED THAT PHYSICIAN WAS UNABLE TO DEFLATE BALLOON EVEN PULLING NEGATIVE WITH INFLATION DEVICE AND SYRINGE. PHYSICIAN WAS FORCED TO PULL OUT THE BALLOON WHILE IT WAS STILL INFLATED AT 6ATM. NO VESSEL DAMAGE OCCURRED. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14111361 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page