Skip to content

RapidCross™ A14BX025150170Recalls & reported problems

Covidien LP · PTA A14BX025150170 RAPIDCROSS V04 · Class 2

All models of RapidCross (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000110857), received between May 14, 2024 and May 14, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2024Deflation Problem; Deflation Problem

    PHYSICIAN ATTEMPTED TO USE A RAPID CROSS PTA BALLOON DILATATION CATHETER WITH A 45CM 6FR NON-MEDTRONIC SHEATH AND A .014 NON-MEDTRONIC GUIDEWIRE DURING TREATMENT OF A FIBROUS LESION IN PATIENTS RIGHT DISTAL ANTERIOR TIBIAL ARTERY (ATA). THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU (INSTRUCTION FOR USE) WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE. EMBOLIC PROTECTION WAS NOT USED.  THE VESSEL WAS PRE-DILATED WITH THE RAPIDCROSS PTA BALLOON. THE VESSEL WAS NOT POST-DILATED. A MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. 50/50 CONTRAST INFLATION FLUID WAS USED. THE BALLOON WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT OF THE DEVICE. DEFLATION DIFFICULTIES WERE REPORTED. NO ISSUES WERE NOTED DURING INFLATION. BALLOON WAS TAKEN TO NOMINAL INFLATION PRESSURE BUT WOULD NOT DEFLATE. THERE WAS NO DAMAGE NOTED TO THE BALLOON CATHETER. THE PHYSICIAN CUT THE HUB OFF THE CATHETER TO ALLOW FOR BALLOON DEFLATION. THE BALLOON WAS EVENTUALLY FULLY DEFLATED FOR REMOV

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19309539 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page