Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000110949), received between Apr 4, 2024 and Apr 4, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 4, 2024Leak/Splash; Device Difficult to Setup or Prepare; Leak/Splash; Device Difficult to Setup or Prepare

    A PHYSICIAN WAS USING A RAPID CROSS PTA BALLOON. IT WAS REPORTED THAT THE  0.014 INCH GUIDEWIRE DOES NOT EXIT THROUGH THE CATHETER GUIDEWIRE PORT LATERALLY, BUT EXITS AT THE PORT CONNECTED TO THE INFLATION LUMEN DISTALLY. WHEN ADMINISTERING SALINE THROUGH THE DISTAL INFLATION PORT, THE OPERATOR NOTICED A LEAK OF SALINE THROUGH THE LATERAL CATHETER GUIDEWIRE PORT. THE OPERATOR DECIDED TO USE ANOTHER BALLOON OF THE SAME REFERENCE (B)(4) TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT. NO DEFORMATION WAS NOTED ON DEVICE. DEVICE WAS USED TO TREAT TIGHT STENOSIS OF THE DISTAL SUPERFICIAL FEMORAL ARTERY. WHEN THE DEVICE WAS RETURNED FOR EVALUATION IT WAS NOTED THAT THERE WAS A LEAK FROM THE TIP OF THE DEVICE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19039718 ↗