MiniMed® 630G MMT-1715KL (UDI 00763000166519)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-1715KL 630G V4.10 BK SF MG · Class 3
All models of MiniMed 630G (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic MiniMed 630G insulin pump has 2,492 adverse event reports, including 10 deaths and 681 injuries. Reports describe cracked or loose reservoir rings, pumps that overheat or will not turn on, and alarm problems, with many patients reporting high blood sugar. Two recalls are open: one over insulin over- or under-delivery when the pump is raised or lowered, and one over parts damaged by drops or bumps that shorten battery life.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,492
- Deaths reported
- 10
- Injuries reported
- 681
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2,492 adverse event reports that list this exact device (DI 00763000166519), received between Jan 2, 2019 and Jun 25, 2025 — classified as 10 deaths, 681 injuries, 1,801 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (714 reports).
2 FDA recall records name this device's identifier, 2 still open.
9 other recalls cover devices cleared under the same FDA submission (P150001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2026Class IIOpen, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1748-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 31, 2024Class IOpen, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0001-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Device Interaction Problem714
- Adverse Event Without Identified Device or Use Problem555
- No Apparent Adverse Event412
- No Display/Image302
- Break245
- Power Problem239
- Connection Problem166
- Obstruction of Flow130
Patient problems
- No Consequences Or Impact To Patient1,327
- Hyperglycemia692
- No Clinical Signs, Symptoms or Conditions306
- Hypoglycemia177
- Elevated Ketones/ Diabetic Ketoacidosis48
- Nausea39
- Vomiting31
- Abdominal Pain16
Recent adverse event reports
The 25 most recent of 2,492 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 25, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED ISSUES WITH THE INSULIN PUMP OVERHEATING. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715KL. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER REPORTED THAT EVEN AFTER CHANGING THE BATTERY 5 TIMES STILL THE PUMP WAS NOT TURNING ON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS REPLACED AND RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22325885 ↗MalfunctionAug 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT UNABLE TO UPLOAD/DOWNLOAD OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145693 ↗MalfunctionJul 28, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS LOOSE AND FALLING OFF. THE LOCATION OF DAMAGE WAS THE RESERVOIR LIP RING. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15120137 ↗MalfunctionJul 25, 2022Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS CRACK AND DAMAGED. THE LOCATION OF DAMAGE WAS THE RESERVOIR RING. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15093694 ↗MalfunctionJul 22, 2022Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15083153 ↗MalfunctionJul 22, 2022Connection Problem; Connection Problem
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15081575 ↗MalfunctionJul 20, 2022Obstruction of Flow; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT NO DELIVERY/OCCLUSION ALARM - UNKNOWN TIMING OCCURRED. THERE WAS AN ALLEGATION OF HYPERGLYCEMIA AS A RESULT OF THIS EVENT.
Patient problems: Hyperglycemia
MAUDE report 15064669 ↗DeathJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT NO DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS AN ALLEGATION OF DEATH AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 15039633 ↗DeathJul 18, 2022Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY ON (B)(6) 2018 IN HOSPITAL. THE CUSTOMER WAS HOSPITALIZED ON (B)(6) 2018 DUE TO HEART ATTACK . THE CAUSE OF DEATH WAS HEART ATTACK. THE CALLER STATED THAT THE CUSTOMER HAD DIABETES THAT MAY HAVE LED TO THE CUSTOMER'S PASSING. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF DEATH. THE CUSTOMER WAS NOT WEARING THE INSULIN PUMP AT THE TIME OF DEATH. THE AUTO MODE ON THE INSULIN PUMP WAS NOT ACTIVE. THE INSULIN PUMP HAD BEEN DISCONNECTED ON (B)(6) 2018 PRIOR TO PASSING. THE CALLER RETURNED THE INSULIN PUMP FOR ANALYSIS.
Patient problems: Myocardial Infarction
MAUDE report 15038464 ↗MalfunctionJul 14, 2022Break; Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS AN ALLEGATION OF REACTION, INJECTION SITE AS A RESULT OF THIS EVENT.
Patient problems: Administration Site Reaction
MAUDE report 15010329 ↗MalfunctionJul 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR HAD AIR BUBBLES. CUSTOMER STATED THAT THEY WERE HOSPITALIZED DUE TO HEART ATTACK. CUSTOMER STATED THAT APPROXIMATE SIZE OF AIR BUBBLES WAS THE SIZE OF THE OPENING OF A PEN. THE RESERVOIR WAS RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 14984735 ↗MalfunctionJul 12, 2022Insufficient Flow or Under Infusion; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Insufficient Flow or Under Infusion; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Hyperglycemia
MAUDE report 14984361 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14955563 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14955132 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954854 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954819 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954689 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS AN ALLEGATION OF HYPERGLYCEMIA AS A RESULT OF THIS EVENT.
Patient problems: Hyperglycemia
MAUDE report 14954268 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954252 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954124 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954107 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954087 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14954069 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14953923 ↗MalfunctionJul 8, 2022Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14953430 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P150001. One submission often covers many models, so check the product and lot details in each record.
- Z-1740-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-1259-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
- Z-0517-2021 · Apr 30, 2020 · Open, ClassifiedMiniMed 630G with SmartGuard, insulin pump, Ref/Model # MMT-1754K, MMT-1714K, MMT-1755K - Product Usage: intended for continuous delivery of basal insulin (at user selectable rate
- Z-0955-2020 · Nov 21, 2019 · Open, ClassifiedMedtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, and MMT-1755 *US and Canada Version Only
- Z-0683-2019 · Oct 3, 2018 · Open, ClassifiedMedtronic MiniMed 630G 6025805-003_C, Rx Only Medtronic MiniMed 670G 6025805-005_B, Rx Only Medtronic MiniMed 640G (International)
- Z-2602-2017 · May 8, 2017 · TerminatedMiniMed 620G Insulin Infusion Pump, Catalog No. MMT-1510, MMT-1710, MMT-1550, MMT-1750.
- Z-2603-2017 · May 8, 2017 · TerminatedMiniMed 630G Insulin Infusion Pump, Catalog No. MMT-1715, MMT-1755.
Other Automated insulin dosing , threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page