Skip to content

Crome™ HF Quad CRT-D MRI SureScan™ DTPC2QQRecalls & reported problems

Medtronic, Inc. · CRTD DTPC2QQ CROME HF QUAD MRI DF4 USA · Class 3 · Implant

All models of Crome HF Quad CRT-D MRI SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Crome HF Quad CRT-D is an implanted heart defibrillator that also delivers pacing therapy to keep the heart in rhythm. The 75 reports mostly describe pocket infections and endocarditis that led to the device being removed, plus leads that came loose from their connector, sensing problems, and batteries running down sooner than expected. Reports describe 1 death and 53 injuries. Four recalls are open, covering a manufacturing weld crack, reduced or no shock energy, reduced shock output, and telemetry errors.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
76
Deaths reported
1
Injuries reported
54
Recalls naming it
4

Summary

The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 00763000178253), received between Nov 14, 2020 and Sep 4, 2026 — classified as 1 death, 54 injuries, 21 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (12 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 4 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 4 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 16, 2024Class IIOpen, Classified

    Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

    Z-2195-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1788-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Jun 22, 2022Class IOpen, Classified

    There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1490-2022 · Root cause: Device Design · Recall details · FDA record ↗

  4. Apr 12, 2022Class IIOpen, Classified

    Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1268-2022 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem33
  • Device Sensing Problem12
  • Pacing Problem10
  • Inappropriate/Inadequate Shock/Stimulation7
  • Premature Discharge of Battery6
  • Appropriate Device Problem Term/Code Not Available5
  • Inappropriate or Unexpected Reset4
  • Electromagnetic Interference2

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Unspecified Infection22
  • Shock from Patient Lead(s)8
  • Erythema4
  • Bradycardia3
  • Discomfort3
  • Endocarditis3
  • Medical device site infection3

Recent adverse event reports

The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Abscess; Fever; Unspecified Infection

    MAUDE report 50618528 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately one month post generator changeout, the patient experienced a pocket infection with Methicillin-resistant Staphylococcus aureus (MRSA). The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Bacterial Infection; Endocarditis; Unspecified Infection

    MAUDE report 50286762 ↗
  • InjuryMay 29, 2026Premature Discharge of Battery

    IT WAS REPORTED BY THE PATIENT'S SPOUSE THAT ALTHOUGH THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS BRAND NEW, IT ONLY HAD A CERTAIN AMOUNT OF BATTERY LIFE REMAINING. THE PATIENT'S SURGEON WAS ALSO REPORTEDLY CONCERNED ABOUT THE ESTIMATED REMAINING LONGEVITY. PLACEMENT OF ONE OF THE PATIENT'S LEADS WAS REPORTEDLY CAUSING THE SITUATION. PROGRAMMING CHANGES WERE ATTEMPTED BUT SETTINGS WERE ULTIMATELY LEFT AS THEY WERE. THE CRT-D AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25329073 ↗
  • InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO THE PATIENT DEVELOPING ENDOCARDITIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY.

    Patient problems: Endocarditis

    MAUDE report 25230642 ↗
  • InjuryMay 15, 2026Misconnection

    IT WAS REPORTED THAT DURING USE OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RIGHT VENTRICULAR (RV) LEAD, IT WAS NOTED THAT THERE WAS A DEVICE CONNECTOR PORT PROBLEM AND THAT THE RV LEAD PIN ENCOUNTERED AN INSERTION ISSUE THAT THE LEAD HAD UNSEATED FROM THE CONNECTOR PORT AND PULLED BACK. THE POCKET WAS RE-OPENED, AND IT WAS FOUND THE RV LEAD WAS NOT FULLY SEATED IN THE HEADER. THE RV LEAD WAS REMOVED AND REINSERTED AND FIXED INTO THE HEADER. THE CRT-D AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25189540 ↗
  • MalfunctionApr 8, 2026Signal Artifact/Noise

    IT WAS REPORTED THAT DURING THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) IMPLANT ATTEMPT, SENSING ONLY NOISE ON THE HEADER IN THE RIGHT VENTRICULAR PORT OCCURRED. THE CRT-D WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820544 ↗
  • InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST POCKET REVISION PROCEDURE AND ANTIBACTERIAL ABSORBABLE ENVELOPE IMPLANT, THE PATIENT EXPERIENCED POCKET EROSION WITH METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS INFECTIVE ENDOCARDITIS. THE PATIENT WAS TREATED WITH ANTIBIOTICS, THE SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Drug Resistant Bacterial Infection; Endocarditis; Pocket Erosion

    MAUDE report 24706432 ↗
  • InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED FOUR DAYS LATER. THE REASON FOR EXPLANT WAS NOT PROVIDED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24622184 ↗
  • InjuryMar 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THREE MONTHS POST-CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM IMPLANT WITH AN ANTIBACTERIAL ABSORBABLE ENVELOPE, THE PATIENT DEVELOPED A POCKET INFECTION. THE PATIENT EXPERIENCED DISCOMFORT, PAIN, REDNESS AND SWELLING. THE CRT-D SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Erythema; Medical device site infection; Pain; Swelling/ Edema

    MAUDE report 24548997 ↗
  • InjuryMar 3, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS FURTHER REPORTED THAT THE THERAPIES WERE CONFIRMED TO BE INAPPROPRIATE. THE PATIENT WAS IN ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS VF. INTERVENTION WAS PERFORMED ON THE CRT-D.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24498854 ↗
  • InjuryFeb 11, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN METHICILLIN-SENSITIVE STAPHYLOCOCCUS AUREUS INFECTION. ADDITIONALLY, THE PATIENT EXPERIENCED BACTEREMIA AND VEGETATION OF THE TRICUSPID VALVE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Bacterial Infection; Medical device site infection

    MAUDE report 24315034 ↗
  • InjuryJan 14, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM EXHIBITED INFECTION. THE SYSTEM WAS EXPLANTED WITHOUT REPLACEMENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 24069793 ↗
  • InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS AFTER IMPLANT OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM, THE PATIENT EXPERIENCED AN INFECTION. THE DEVICE AND LEADS WERE WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 23586155 ↗
  • InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE RIGHT VENTRICULAR (RV) LEAD AND LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 23586011 ↗
  • InjuryOct 24, 2025Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND DRAINED THE BATTERY OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SUBSTANTIALLY. THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383662 ↗
  • InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS REVISED DUE TO EROSION. THE DEVICE WAS INITIALLY IMPLANTED SUBCUTANEOUS AND WAS REVISED TO BE PLACED SUBMUSCULAR. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Erosion; Pocket Erosion

    MAUDE report 22680530 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY 4 MONTHS POST IMPLANT PROCEDURE, THE PATIENT DEVELOPED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 22301688 ↗
  • MalfunctionJun 18, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE PATIENT'S REPRESENTATIVE THAT THE PATIENT WAS TO REPORT TO THE EMERGENCY DEPARTMENT AFTER HEARING THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) TONING TWO DAYS IN A ROW IN THE AFTERNOON. THE PATIENT'S REPRESENTATIVE NOTED THAT THE CLINICIANS AT THE HOSPITAL EXAMINED THE DATA AND THERE WAS NOTHING OBSERVED. THE CLINICIAN NOTED THAT "NOTHING TRIGGERED". THE PATIENT'S REPRESENTATIVE WAS CONCERNED AS TO WHY THE CRT-D WOULD EMIT A HIGH/LOW TONE BUT NOT MAKE ANY OBSERVATIONS AS TO WHY IT WAS ALERTING. THE REPRESENTATIVE WAS CONCERNED "SOMETHING MAY NOT BE WORKING PROPERLY". THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286962 ↗
  • MalfunctionMay 30, 2025Electromagnetic Interference; Device-Device Incompatibility; Electromagnetic Interference; Device-Device Incompatibility

    IT WAS REPORTED THAT MULTIPLE LEAD INTEGRITY ALERTS (LIA) TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD DUE TO HIGH RATE NON-SUSTAINED (HRNS) EPISODES AND SHORT VENTRICULAR INTERVALS. SHORT RUNS OF OVERSENSING WERE NOTED AND APPEARED CONSISTENT OVER THE COURSE OF THE LIAS. INTERFERENCE WAS REPORTEDLY NOTED WITH THE PATIENT'S CARDIAC CONTRACTILITY MODULATION (CCM) DEVICE WHICH WAS SUBSEQUENTLY DEACTIVATED; HOWEVER, THE HRNS EPISODES WERE NOTED TO HAVE OCCURRED PRIOR TO CCM IMPLANT AND POST-DEACTIVATION OF THE DEVICE. STABLE IMPEDANCE AND SENSING TRENDS AND THE ABSENCE OF A RISE IN THE FREQUENCY OF SHORT VENTRICULAR INTERVALS SUGGESTED AN ELECTROMAGNETIC INTERFERENCE (EMI) SOURCE OR MYOPOTENTIALS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22112972 ↗
  • InjuryApr 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO ARRHYTHMIA NOTED AT THEIR FOLLOW-UP VISIT. THE PATIENT WAS DISCHARGED HOME AND THEN FELT THE DEVICE VIBRATING. THE PATIENT HAD ALSO BEEN EXPERIENCING TIREDNESS. THE PATIENT WAS ON HOSPICE FOR EXTRA CARE WHILE WAITING FOR A FUTURE LEFT VENTRICULAR ASSIST DEVICE (LVAD). THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Arrhythmia; Fatigue

    MAUDE report 21881552 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT MORE THAN A MONTH AFTER IMPLANT, THE PATIENT EXPERIENCED AN INFECTION WITH SWELLING, PURULENT DISCHARGE/PUS, PAIN AND DISCOMFORT. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Pain; Purulent Discharge; Swelling/ Edema; Unspecified Infection

    MAUDE report 21652022 ↗
  • InjuryMar 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), RIGHT ATRIAL (RA) AND LEFT VENTRICULAR (LV) LEADS WERE EXPLANTED AND THE RIGHT VENTRICULAR (RV) LEAD WAS CAPPED DUE TO A POCKET INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection

    MAUDE report 21519001 ↗
  • MalfunctionJan 23, 2025Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT ATRIAL EVENTS APPEARED TO BE FALLING IN BLANKING AT TIMES POSSIBLY TRIGGERING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) DEVICE CLASSIFIED TERMINATION OF AT/AF DETECTED EVENT(S) AND OCCASIONAL INAPPROPRIATE ATRIAL PACING. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21225873 ↗
  • InjuryJan 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST IMPLANT PROCEDURE, THE PATIENT DEVELOPED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 21208424 ↗
  • InjuryJan 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION DUE TO HIGH LEFT VENTRICULAR (LV) LEAD THRESHOLDS. DURING THE LV LEAD EXTRACTION, THE SUPERIOR VENA CAVA (SVC) DISSECTED WHICH WAS NOTED TO BE CONTAINED. THE LV LEAD AND CRT-D WERE EXPLANTED AND REPLACED. AFTER CONNECTING THE LEADS TO THE NEW DEVICE AND PERFORMING AN AUTOMATED PACING POLARITIES TEST, THE PATIENT WENT INTO VENTRICULAR FIBRILLATION (VF) AND REQUIRED TO BE EXTERNALLY SHOCKED, WHICH WAS SUCCESSFUL. THE NEW CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Ventricular Fibrillation

    MAUDE report 21106404 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page