Skip to content

EverCross™ AB35W04120135 (UDI 00763000184018)Recalls & reported problems

Ev3, Inc · PTA AB35W04120135 EVERCROSS 035 V09 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000184018), received between Feb 9, 2021 and Feb 9, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2021Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN INTENDED TO USE A EVERCROSS FOR PATIENT TREATMENT ON THE SUPERFICIAL FEMORAL ARTERY (SFA). THE MID LESION WAS SEVERELY CALCIFIED AND NO TORTUOSITY. CHRONIC TOTAL OCCLUSION OF 100% LESION STENOSIS REPORTED. THERE WAS NO DAMAGE NOTED TO THE PACKAGING NOR ANY ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. IFU WAS FOLLOWED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE BALLOON WAS INFLATED WITH AN INFLATION DEVICE. A BALLOON BURST WAS REPORTED TO OCCUR DURING BALLOON INFLATION AT 8 ATM. THE BALLOON DID NOT FRAGMENT. NO INTERVENTION WAS REQUIRED FOR REMOVAL OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH A BALLOON OF THE SAME SIZE, USED TO BALLOON AND PLACE A STENT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11299974 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates