Skip to content

EverCross™ AB35W05020135Recalls & reported problems

Ev3, Inc · PTA AB35W05020135 EVERCROSS 035 V09 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000184087), received between Oct 7, 2019 and Oct 7, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Detachment of Device or Device Component1

Patient problems

  • Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 7, 2019Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component

    PHYSICIAN USED EVERCROSS PTA BALLOON CATHETER ALONG WITH NON MEDTRONIC 7FR SHEATH AND .035 GUIDE WIRE TO TREAT A NONE CALCIFIED SOFT TISSUE LESION IN THE RIGHT PROXIMAL ILIAC COMMON ARTERY. THE VESSEL DIAMETER WAS 8MM AND SEVERELY TORTUOUS. NEGATIVE PREP WAS PERFORMED. A NON-MEDTRONIC INFLATION DEVICE WAS USED WITH INFLATION FLUID. THERE WAS NO DAMAGE NOTED TO PACKAGING AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED PER THE IFU WITH NO ISSUES IDENTIFIED. IT WAS REPORTED THAT THE BALLOON WAS USED TO TREAT THE AORTIC BIFURCATION TO HELP CROSSOVER KISSING STENTS. THE BALLOON WAS INFLATED TO 6-7ATM AND IT BURST. THE BALLOON IS THOUGHT TO HAVE GONE THROUGH A STENT STRUT AND UPON REMOVAL WAS CAUGHT ON STRUT. MAKE AND MODEL OF PREVIOUSLY DEPLOYED STENT IS UNKNOWN. THE BALLOON TIP WAS DETACHED DURING REMOVAL AND SNARE USED TO REMOVE DETACHED COMPONENT.

    Patient problems: Injury

    MAUDE report 9162008 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page