EverCross™ AB35W05040135 (UDI 00763000184148)Recalls & reported problems
Ev3, Inc · PTA AB35W05040135 EVERCROSS 035 V09 · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000184148), received between Aug 23, 2019 and Feb 3, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Burst Container or Vessel2
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 3, 2023Burst Container or Vessel; Burst Container or Vessel
AN EVERCROSS PTA BALLOON WAS BEING USED DURING THE TREATMENT OF A LESION IN THE SUPERFICIAL FEMORAL ARTERY. NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. NO DAMAGE NOTED TO PACKAGING. NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT OF THE DEVICE. IT WAS REPORTED THE BALLOON RUPTURED IN THE PATIENT AND WAS WITHDRAWN. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16290873 ↗MalfunctionNov 12, 2021Leak/Splash; Leak/Splash
PHYSICIAN ATTEMPTED TO USE AN EVERCROSS PTA BALLOON DURING TREATMENT OF A CALCIFIED LESION IN THE PATIENT'S MID SUPERFICIAL FEMORAL ARTERY (SFA). SLIGHT VESSEL CALCIFICATION IS REPORTED. LESION EXHIBITED 80% STENOSIS. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. DEVICE PREPPED AS PER IFU WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT AND NO EXCESSIVE FORCE WAS USED. THE BALLOON WAS FOUND TO BE LEAKING DURING USE AND CANNOT BE USED MULTIPLE TIMES. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12802046 ↗MalfunctionAug 23, 2019Burst Container or Vessel; Burst Container or Vessel
PHYSICIAN USED AN EVERCROSS BALLOON WITH A 6FR SHEATH AND 0.011" GUIDEWIRE DURING TREATMENT OF A 100MM PLAQUE LESION IN THE PATIENT¿S MID SUPERFICIAL FEMORAL ARTERY (SFA) OF DIAMETER 4MM. SLIGHT TORTUOSITY AND MODERATE CALCIFICATION ARE REPORTED. LESION EXHIBITED 90% STENOSIS. IFU WAS FOLLOWED, DEVICE PREPPED WITHOUT ISSUE. THE DEVICE WAS PASSED THROUGH A PREVIOUSLY DEPLOYED EVERFLEX ENTRUST STENT WITHOUT RESISTANCE. AN INFLATION DEVICE WAS USED WITH 50/50 INFLATION FLUID. IT IS REPORTED DURING BALLOON INFLATION AT A PRESSURE OF 8ATM A PIN HOLE BURST OCCURRED. THIS OCCURRED ON SECOND INFLATION OF THE BALLOON IN AN INTRASTENT STENT. NO FRAGMENTS OF THE BALLOON REMAINED IN THE PATIENT. AN 8X60 EVERCROSS BALLOON WAS THEN USED TO COMPLETE THE PROCEDURE. NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8924962 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.