Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000184346), received between Nov 1, 2021 and Nov 1, 2021 — classified as 1 death.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • Myocardial Infarction1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathNov 1, 2021Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN ATTEMPTED TO USE AN EVERCROSS PTA BALLOON DURING TREATMENT OF A CALCIFIED LESION IN THE PATIENT'S LEFT COMMON ILIAC ARTERY. SEVERE VESSEL CALCIFICATION IS REPORTED. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED AS PER IFU WITHOUT ISSUE. INFLATION DEVICE WAS USED FOR BALLOON INFLATION. A BALLOON BURST IS REPORTED. THE DEVICE HAD TO BE SURGICALLY REMOVED FROM THE PATIENT'S ARTERY. THE PATIENT IS NOW DECEASED, THE PATIENT HAD A MYOCARDIAL INFARCTION DAYS AFTER THE CASE.

    Patient problems: Myocardial Infarction

    MAUDE report 12726450 ↗