EverCross™ AB35W06120135 (UDI 00763000184476)Recalls & reported problems

Ev3, Inc · PTA AB35W06120135 EVERCROSS 035 V09 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000184476), received between Mar 9, 2021 and Jun 21, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2
  • Deflation Problem1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2024Burst Container or Vessel; Deflation Problem; Burst Container or Vessel; Deflation Problem

    PHYSICIAN INTENDED TO USE AN EVERCROSS 035 PTA BALLOON WITH A 6FR SHEATH AND A 0.035 GUIDEWIRE DURING TREATMENT OF A 120MM FIBROUS LESION IN THE PATIENTS RIGHT MID DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). MODERATE VESSEL TORTUOSITY AND MODERATE VESSEL CALCIFICATION WERE REPORTED. LESION EXHIBITED 70% STENOSIS. ARTERY DIAMETER REPORTED AS 5MM. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED PER THE IFU (INSTRUCTION FOR USE) WITH NO ISSUES IDENTIFIED. EMBOLIC PROTECTION WAS NOT USED. AN INFLATION DEVICE WAS USED WITH NO ISSUES NOTED DURING INFLATION. CONTRAST INFLATION FLUID WAS USED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE AND THE BALLOON DID NOT DETACH DURING THE EVENT. THE DEVICE WAS SLOW TO DEFLATE AT THE LESION SITE. DEFLATION TOOK 40 SECONDS. A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCC URRED AT 7 ATMS. NO VESSEL INJURY NOTED. A REPLACEMENT NON-MEDTRONIC BALLOON OF THE SAME SIZE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJU

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19581506 ↗
  • InjuryMar 9, 2021Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove

    PHYSICIAN WAS ATTEMPTING TO USE AN EVERCROSS PTA BALLOON DURING PROCEDURE TO TREAT A SEVERELY CALCIFIED LESION IN THE SUPERFICIAL FE MORAL ARTERY (SFA). THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUE NOTED WHEN REMOVING THE DEVICE FROM HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. THERE WAS DIFFICULTY REMOVING BALLOON FOLLOWING BALLOON INFLATION. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THERE WAS RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE. IT WAS REPORTED THAT BALLOON GOT RUPTURED IN SFA, HALF OF THE BALLOON GOT STUCK DUE TO THE CALCIUM IN THE VESSEL, TWO PHYSICIANS TRIED TO SNARE IT OUT, BUT WAS UNSUCCESSFUL. FINALLY CONTRAST RUN WAS DONE AND PATIENT HAD GOOD FLOW, FOR THAT REASON, PHYSICIAN DID NOT BYPASS THE PATIENT. PATIENT WAS SENT TO ANOTHER HOSPITAL TO KEEP AN EYE ON AS NO VASCULAR SURGEON WAS ON CALL AT THE TREATMENT FACILITY. PHYSICIAN DID NOT DO ANY DATIX ON THIS INCIDENT, AS THIS WAS NOT AN AVOIDABLE INCIDENT DUE TO CALCIUM, BALLOON GOT STUCK. NO FURTHER PATIENT INJURY REPORTED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11445791 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →