EverCross™ AB35W08040135 (UDI 00763000184803)Recalls & reported problems
Ev3, Inc · PTA AB35W08040135 EVERCROSS 035 V09 · Class 2
All models of EverCross (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000184803), received between Dec 26, 2019 and Nov 26, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2021Burst Container or Vessel; Burst Container or Vessel
PHYSICIAN WAS ATTEMPTING TO USE THE EVERCROSS PTA DURING ALONG WITH 5.0 SPIDER FX EMBOLIC PROTECTION DURING PROCEDURE TO TREAT A MODERATELY CALCIFIED LESION IN THE MID TO DISTAL POPLITEAL ARTERY (POP) WITH 40% STENOSIS. THE VESSEL WAS NONE TORTUOUS. A MEDTRONIC INFLATION DEVICE WAS USED FOR INFLATION. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUE NOTED WHEN REMOVING DEVICE FROM HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. IT WAS REPORTED THAT DURING BALLOON INFLATION, A BURST OCCURRED AND ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PTA BALLOON DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THERE WAS NO RESISTANCE ENCOUNTERED WHEN ADVANCING DEVICE. PROCEDURE WAS COMPLETED AFTER BALLOON BURST. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12877936 ↗MalfunctionDec 26, 2019Burst Container or Vessel; Burst Container or Vessel
PHYSICIAN INTENDED TO USE AN EVERCROSS PTA BALLOON DURING TREATMENT OF A PLAQUE LESION IN THE PATIENTS¿ MID SUPERFICIAL FEMORAL ARTERY (SFA). SLIGHT VESSEL TORTUOSITY AND CALCIFICATION ARE REPORTED. LESION EXHIBITED 60% STENOSIS. A NON-MEDTRONIC INFLATION DEVICE WAS USED. NO DAMAGE NOTED TO PACKAGING PRIOR TO USE. NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED. THE DEVICE WAS PREPPED WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. IT IS REPORTED THAT A LONGITUDINAL BALLOON BURST OCCURRED DURING BALLOON INFLATION AT A PRESSURE OF 8ATM. NO BALLOON FRAGMENTS WERE LEFT IN THE PATIENT. THE DEVICE WAS REPLACED WITH ANOTHER MEDTRONIC DEVICE TO COMPLETE THE PROCEDURE. NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9519252 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page