Skip to content

EverCross™ AB35W08060135 (UDI 00763000184834)Recalls & reported problems

Ev3, Inc · PTA AB35W08060135 EVERCROSS 035 V09 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000184834), received between Jul 2, 2024 and Apr 25, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Delivered as Unsterile Product1
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT PRIOR TO USE OF THE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY AB35W08060135 EVERCROSS 035 BALLOON, THERE WERE PACKAGING ISSUES. THE OUTER PACKAGE WAS DAMAGED, AND THERE WAS A TEAR ON THE PACKAGE, WHICH COMPROMISED THE STERILE PACKAGING. THE PROCEDURE INTENDED TO TREAT A CALCIFIED LESION IN THE RIGHT MID COMMON ILIAC ARTERY. THE DEVICE WAS NOT USED IN THE PATIENT, AND A NEW EVERCROSS 8X60X80 DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21907761 ↗
  • InjuryJul 2, 2024Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component

    PHYSICIAN ATTEMPTED TO USE AN EVERCROSS 035 PTA BALLOON WITH A 6FR SHEATH AND A .014 GUIDEWIRE DURING TREATMENT OF A 60MM FIBROUS LESION IN THE PATIENTS RIGHT MID COMMON ILIAC ARTERY. NO VESSEL TORTUOSITY AND MINIMAL VESSEL CALCIFICATION WERE REPORTED. ARTERY DIAMETER REPORTED AS 7.5MM. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. THE DEVICE WAS PREPPED PER THE IFU (INSTRUCTION FOR USE) WITH NO ISSUES IDENTIFIED. EMBOLIC PROTECTION WAS NOT USED. AN INFLATION DEVICE WAS USED FOR BALLOON INFLATION. CONTRAST INFLATION FLUID WAS USED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCCURRED ON FIRST INFLATION AT 8ATMS. THE BALLOON BROKEN HORIZONTALLY AND HALF OF THE BALLOON WAS IN THE PATIENT'S BODY. DIFFICULTY REMOVING BALLOON FOLLOWING BALLOON INFLATION WAS ALSO REPORTED. PHYSICIAN MADE A CUT ON SKIN FOR RETRIEVAL OF THE DETACHED COMPONENT USING A SNARE. A REPLACEMENT NON-MEDTRONIC BALLOON WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19655429 ↗

Other Catheter, angioplasty, peripheral, transluminal devices