Affinity Fusion™ CB811 (UDI 00763000225476)Recalls & reported problems
Medtronic, Inc. · FPA FUSION OXY CB811 CORTIVA 2L · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000225476), received between Nov 29, 2022 and Mar 20, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K183490, K203111); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1025-2023 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Material Separation2
- Gas Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, THE CUSTOMER REPORTED THAT THE DEVICE WAS SET UP AND WATER TESTED. UPON TURNING THE HEATER/COOLER ON IN ORDER TO WATER TEST THE DEVICE, THE CUSTOMER NOTED WATER SPRAYING OUT OF THE GAS EXHAUST PORT. THE DEVICE WAS REPLACED IMMEDIATELY TO RUN THE CASE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE OR PACK/PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18938846 ↗MalfunctionNov 28, 2023Fluid/Blood Leak; Gas Output Problem; Fluid/Blood Leak; Gas Output Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CARMEDA FUSION HOLLOW FIBER OXYGENATOR, IT WAS REPORTED THAT DURING SURGERY WITH ECC, THE DEVICE WORKED POORLY. CO2 WAS DELIVERED TO THE OPEN CHEST, AND THE OXYGEN DELIVERY WAS GOOD. BUT THEN THE CARBON DIOXIDE VENTILATION BEGAN TO DETERIORATE SIGNIFICANTLY FROM THE NORMAL SPEED OF 3-4 LITERS OF FRESH GAS FLOW. 13 LITERS ARE REQUIRED TO KEEP PCO2 BELOW 6KPA. THE DEVICE COULD ONLY DELIVER 10 LITERS, SO AN EXTRA FRESH GAS HOSE WAS SPLICED ONTO THE EXISTING ONE, AND 3 LITERS WAS THEN SUPPLIED FROM THE ANESTHESIA MACHINE'S ROTAMETER. ANESTHETIC GAS ALSO LEAKED FROM THE OXYGENATOR, AND THIS FORCED CUSTOMER TO CHANGE THE ANESTHESIA STRATEGY TO PROPOFOL INFUSION. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18215055 ↗MalfunctionJun 15, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE ARTERIAL TEMPERATURE PROBE FOR THE OXYGENATOR CAME LOOSE WHEN DISASSEMBLING THE DISPOSABLE PRODUCTS FROM THE HEART LUNG MACHINE POST-OP. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17136573 ↗MalfunctionNov 29, 2022Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION HOLLOW FIBER OXYGENATOR, THE CUSTOMER REPORTED THAT THE TEMPERATURE PORT ON THE DEVICE CAME OUT COMPLETELY FROM THE MOLDING WHEN UNSCREWING THE PROBE FROM THE TEMP PORT. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT SINCE THE CARDIOPULMONARY BYPASS PART OF THE CASE WAS COMPLETELY OVER. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE PATIENT WAS NOT CONNECTED TO THE PERFUSION PUMP IN ANY WAY WHEN THE TEMPERATURE PORT BECAME LOOSE AND CAME OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15876368 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K183490, K203111. One submission often covers many models, so check the product and lot details in each record.
- Z-1026-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass.
- Z-1028-2023 · Dec 12, 2022 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF CB811. Used in Cardiopulmonary bypass.
- Z-1023-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF BB811. Used in Cardiopulmonary bypass.
- Z-1027-2023 · Dec 12, 2022 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF BB811. Used in Cardiopulmonary bypass.
- Z-1024-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, OXY/CVR COMBO BB841 FUSION W/BALANCE 26L, REF BB841. Used in Cardiopulmonary bypass.
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-iD QUADROX-iD Adult, SOFTLINE Coating, without FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports