Skip to content

MiniMed® 670G MMT-1780KL (UDI 00763000256821)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-1780KL 670G V4.11 JBL MG · Class 3

All models of MiniMed 670G (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This MiniMed 670G insulin pump has 374 reports, with no deaths and 93 injuries. Many describe power problems, cracked or broken parts, and blank screens, and some mention high or low blood sugar. There are 2 recalls, both still open: a 2026 one over insulin over- or under-delivery when the pump is raised or lowered, and a 2024 one over pumps that were dropped or bumped and may lose battery life.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
374
Deaths reported
0
Injuries reported
93
Recalls naming it
2

Summary

The FDA's MAUDE database contains 374 adverse event reports that list this exact device (DI 00763000256821), received between Sep 5, 2019 and Jul 27, 2022 — classified as 93 injuries, 281 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (131 reports).

2 FDA recall records name this device's identifier, 2 still open.

12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Feb 13, 2026Open, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 31, 2024Open, Classified

    Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

    Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Patient Device Interaction Problem131
  • Adverse Event Without Identified Device or Use Problem85
  • No Apparent Adverse Event65
  • Power Problem50
  • Break45
  • No Display/Image28
  • Mechanical Problem24
  • Connection Problem19

Patient problems

  • No Consequences Or Impact To Patient202
  • Hyperglycemia80
  • No Clinical Signs, Symptoms or Conditions60
  • Hypoglycemia34
  • Elevated Ketones/ Diabetic Ketoacidosis9
  • Nausea6
  • Abdominal Pain4
  • Vomiting4

Recent adverse event reports

The 25 most recent of 374 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2022Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD CRACKED RETAINER RING AND THE RESERVOIR WAS ABLE LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15113648 ↗
  • MalfunctionJul 8, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14954573 ↗
  • MalfunctionJul 8, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14954333 ↗
  • MalfunctionJul 8, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14953770 ↗
  • MalfunctionJul 8, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14952944 ↗
  • MalfunctionJul 4, 2022No Display/Image; Electrical /Electronic Property Problem; Device Markings/Labelling Problem

    MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, LABELING ANOMALY, PUMP ERROR 23 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14906174 ↗
  • MalfunctionJul 4, 2022Power Problem; Intermittent Loss of Power

    MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 82, UNEXPECTED BATTERY POWER LOSS OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14905788 ↗
  • MalfunctionJul 4, 2022Electrical /Electronic Property Problem

    MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 23 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14905082 ↗
  • MalfunctionJul 4, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14904577 ↗
  • MalfunctionJul 4, 2022Circuit Failure; Electrical /Electronic Property Problem; Communication or Transmission Problem; Intermittent Communication Failure

    MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 24, PUMP ERROR 3, PUMP ERROR 81, PUMP ERROR 23 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14904472 ↗
  • MalfunctionJun 29, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS AN ALLEGATION OF HYPERGLYCEMIA AS A RESULT OF THIS EVENT.

    Patient problems: Hyperglycemia

    MAUDE report 14865285 ↗
  • MalfunctionJun 26, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14823872 ↗
  • MalfunctionJun 25, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14816170 ↗
  • MalfunctionJun 25, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14816080 ↗
  • MalfunctionJun 25, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14813855 ↗
  • MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14796444 ↗
  • MalfunctionJun 24, 2022Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14796376 ↗
  • MalfunctionJun 23, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14795673 ↗
  • MalfunctionJun 23, 2022Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14795565 ↗
  • MalfunctionJun 23, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14795227 ↗
  • MalfunctionJun 23, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14785741 ↗
  • MalfunctionJun 23, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14785079 ↗
  • MalfunctionJun 23, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14784787 ↗
  • MalfunctionJun 22, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14782382 ↗
  • MalfunctionJun 22, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14781654 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2981-2026 · Jul 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
  • Z-1742-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1261-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1265-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1266-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
  • Z-0193-2023 · Sep 20, 2022 · Open, Classified
    Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
  • Z-0855-2022 · Jan 27, 2022 · Open, Classified
    MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-18
  • Z-1771-2021 · Mar 15, 2021 · Open, Classified
    MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE
  • Z-0516-2021 · Apr 30, 2020 · Open, Classified
    MiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
  • Z-0956-2020 · Nov 21, 2019 · Open, Classified
    Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
  • Z-2604-2017 · May 8, 2017 · Terminated
    MiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
  • Z-2605-2017 · May 8, 2017 · Terminated
    MiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.

Other Automated insulin dosing device system, single hormonal control devices

All Automated insulin dosing device system, single hormonal control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page