MiniMed® 670G MMT-1760KPK (UDI 00763000283445)Recalls & reported problems

Medtronic Minimed, Inc. · KIT MMT-1760KPK 670G V4.11 MG · Class 3

All models of MiniMed 670G (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is the MiniMed 670G insulin pump kit. The 38 reports mostly describe pumps losing power, insulin flow being blocked, cracked battery caps, and blank or unresponsive screens, alongside blood sugar swings in some users. There are 16 injuries and no deaths. Three open recalls cover unintended over- or under-delivery of insulin, shortened battery life after drops or bumps, and battery cap deterioration that can cut power.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
16
Recalls naming it
3

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 00763000283445), received between Mar 4, 2020 and Jun 24, 2022 — classified as 16 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Obstruction of Flow” (5 reports).

3 FDA recall records name this device's identifier, 3 still open.

12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 31, 2024Class IOpen, Classified

    Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

    Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. May 23, 2022Class IIOpen, Classified

    Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.

    Z-1355-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem12
  • No Apparent Adverse Event5
  • Obstruction of Flow5
  • Device Difficult to Program or Calibrate4
  • Power Problem4
  • Mechanical Problem3
  • No Display/Image3
  • Break2

Patient problems

  • No Consequences Or Impact To Patient16
  • Hyperglycemia13
  • Hypoglycemia8
  • No Clinical Signs, Symptoms or Conditions3
  • Abdominal Pain1
  • Headache1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2022Battery Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, LOW BATTERY ALERT LESS THAN 1 WEEK OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14797153 ↗
  • MalfunctionSep 17, 2021Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING OF THE INSULIN PUMP IS MISSING. RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12488298 ↗
  • MalfunctionSep 17, 2021Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING OF THE INSULIN PUMP IS MISSING. RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 12488174 ↗
  • InjuryAug 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT CUSTOMER WERE EXPERIENCING LOW BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 51 MG/DL. CUSTOMER TREATED WITH FOOD. CUSTOMER STATED AUTO MODE FEATURE WAS NOT ACTIVE AT THE TIME OF INCIDENT. CUSTOMER DID NOT WANT TO TROUBLESHOOTING FOR LOW BLOOD GLUCOSE. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10389769 ↗
  • MalfunctionAug 3, 2020Use of Device Problem; Power Problem; Use of Device Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP POWER LOSS ALARM. CUSTOMER WAS UNABLE TO EXIT TO LOAD RESERVOIR PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10355697 ↗
  • MalfunctionJul 31, 2020No Display/Image; Mechanical Problem; Moisture or Humidity Problem; No Display/Image; Mechanical Problem; Moisture or Humidity Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AND BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY ERROR. THE INSULIN PUMP WAS EXPOSED TO MOISTURE. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND THE DISPLAY DID NOT RETURN. CUSTOMER STATED THAT THE DISPLAY WAS NOT BLANK LESS THAN 30 SECONDS AND DID NOT RETURN. CUSTOMER STATED THAT THE CONTACTS ON THE BATTERY CAP WERE NOT MISSING OR DAMAGED. CUSTOMER STATED THAT THE BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED ALSO THE SPRING WAS NEITHER DAMAGED NOR CORRODED. THE DISPLAY DID NOT RETURN AFTER INSULIN PUMP RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10348645 ↗
  • InjuryJul 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 400 MG/DL. THE CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER REPORTED THAT THEY TREATED HIGH BLOOD GLUCOSE LEVEL WITH INSULIN PUMP BOLUS. CUSTOMER WAS NEITHER IN EMERGENCY ROOM, NOR ADMITTED INTO HOSPITAL, NOR WAS EMERGENCY MEDICAL SERVICES DISPATCHED AS A RESULT OF HIGH BLOOD GLUCOSE. THE CUSTOMER DID NOT ALLEGE THE INSULIN PUMP FOR UNDER DELIVERY. THE INSULIN PUMP WAS ON AUTO MODE AT THE TIME OF THE INCIDENT. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10337517 ↗
  • InjuryJul 29, 2020Obstruction of Flow; Obstruction of Flow

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAD HYPERGLYCEMIA. BLOOD GLUCOSE LEVEL WAS 500 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP GOT INSULIN FLOW BLOCK ALARM. CUSTOMER ALSO MENTIONED THAT THE RESERVOIR WAS EMPTY. CUSTOMER DECLINED TO TROUBLESHOOTING FOR THE HIGH BLOOD GLUCOSE. CUSTOMER STATED THAT EVERY THREE DAYS TRANSMITTER GOT DISCONNECTED AND WHEN CUSTOMER RECONNECT IT DID NOT SEND INFORMATION TO THE INSULIN PUMP. IT WAS UNKNOWN WHETHER THEY USING AUTO MODE FEATURE AT THE TIME OF REPORTED HIGH BLOOD GLUCOSE INCIDENT. INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10336698 ↗
  • InjuryJul 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT CUSTOMER WERE EXPERIENCING HIGH BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 499 MG/DL. CUSTOMER WAS USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. CUSTOMER TREATED WITH BLOUSING WITH INSULIN PUMP. CUSTOMER WAS NEITHER IN EMERGENCY ROOM NOR ADMITTED INTO HOSPITAL. BASED ON CUSTOMER REPORT CUSTOMER DID NOT ALLEGE INSULIN PUMP WAS UNDER DELIVERING. CUSTOMER STATED AUTO MODE FEATURE WAS ACTIVE AT THE TIME OF INCIDENT. CUSTOMER DECLINED TO TROUBLESHOOTING THE SENSOR UPDATING INCIDENT. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10325606 ↗
  • MalfunctionJul 13, 2020Electrical /Electronic Property Problem; Device Difficult to Program or Calibrate; Loss of Data; Key or Button Unresponsive/not Working; Electrical /Electronic Property Problem; Device Difficult to Program or Calibrate; Loss of Data; Key or Button Unresponsive/not Working

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MULTIPLE INSULIN PUMP ERROR ALARMS. THE CUSTOMER STATED THAT INSULIN PUMP DISPLAY WAS BLANK AND FROZEN. CUSTOMER WAS NOT ABLE TO CLEAR THE ALARM. BUTTONS WERE NOT RESPONDING AS WELL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10263287 ↗
  • MalfunctionJul 10, 2020Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED CRITICAL PUMP ERROR. CUSTOMER REPORTED THAT THEY RECEIVED OPEN BOOK ERROR ON THE INSULIN PUMP SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10256349 ↗
  • MalfunctionJul 3, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED WITH AN INSULIN PUMP ERROR AND STARTED BEEPING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10229285 ↗
  • MalfunctionJun 30, 2020Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A CRITICAL PUMP ERROR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. CUSTOMER REPORTED THAT THEY RECEIVED THE OPEN BOOK IMAGE ON THE INSULIN PUMP SCREEN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10210963 ↗
  • InjuryJun 25, 2020Obstruction of Flow; Obstruction of Flow

    CUSTOMER CALLED AND REPORTED THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE WITH BLOOD GLUCOSE OF 485 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER DID NOT MENTION HOW THEY TREATED. THE CUSTOMER DID NOT MENTION ANY SYMPTOMS. THE CUSTOMER WAS WEARING THE INSULIN PUMP DURING THE INCIDENT. THE AUTO MODE ON THE INSULIN PUMP WAS NOT ACTIVE AND AUTOMATICALLY ADJUSTING INSULIN DELIVERY DURING THE EVENT. TROUBLESHOOTING WAS NOT COMPLETED AS THE CUSTOMER DECLINED. THE CUSTOMER WAS GETTING INSULIN FLOW BLOCKED ALARMS AND STOPPED USING THE PUMP. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10198157 ↗
  • InjuryJun 25, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 450 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER DID NOT EXPERIENCED SYMPTOMS DUE TO HIGH BLOOD GLUCOSE. THE CUSTOMER TREATED HIGH BLOOD GLUCOSE WITH BOLUS. THE INSULIN PUMP WAS ON AUTO MODE AT THE TIME OF INCIDENT. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER DID NOT ALLEGE UNDER DELIVERY ON INSULIN PUMP. THE CUSTOMER WAS NEITHER IN THE EMERGENCY ROOM, NOR ADMITTED INTO HOSPITAL AS A RESULT OF HIGH BLOOD GLUCOSE. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10193726 ↗
  • MalfunctionJun 24, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Power Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY AND CRITICAL PUMP ERROR ALARM. CUSTOMER STATED BUTTONS DON'T WORK, IT ASKED TO BE REWOUND AND GOT A CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10187895 ↗
  • InjuryJun 23, 2020Device Alarm System; Device Alarm System

    IT WAS REPORTED THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL OF 430 MG/DL. INSULIN WAS LEAKING FROM TUBING. THE CUSTOMER DID PROVIDE SYMPTOMS REGARDING HIGH BLOOD GLUCOSE SUCH AS ABDOMINAL PAIN AND HEADACHE. CUSTOMER REPORTED THAT THE INSULIN PUMP ALSO RECEIVED INSULIN FLOW BLOCK ALARM. THE CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. CUSTOMER WAS NOT USING AUTO MODE FEATURE AT THE TIME OF REPORTED EVENT. THE INSULIN PUMP WILL NOT RETURN FOR ANALYSIS.

    Patient problems: Abdominal Pain; Headache; Hyperglycemia

    MAUDE report 10183121 ↗
  • MalfunctionJun 19, 2020Device Difficult to Program or Calibrate; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTONS NOT WORK. THE CUSTOMER STATED THAT THE CRACKED ON HOUSING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10172747 ↗
  • InjuryJun 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CUSTOMER HAD LOW BLOOD GLUCOSE OF 50 MG/DL. CUSTOMER TREATED WITH FOOD FOR THEIR LOW BLOOD GLUCOSE. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED LOW BLOOD GLUCOSE EVENT. CUSTOMER DID NOT ALLEGE INSULIN PUMP FOR OVER DELIVERY. CUSTOMER WAS USING AUTO MODE AT THE TIME OF REPORTED LOW BLOOD GLUCOSE EVENT. INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10161403 ↗
  • MalfunctionJun 15, 2020No Display/Image; No Display/Image

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DISPLAY DID NOT RETURNED BACK. THE CUSTOMER WAS ALSO REPORTED THAT BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED, BATTERY CAP WAS NEITHER MISSING NOR DAMAGED, AND THE SPRING WAS NEITHER DAMAGED NOR CORRODED. THE CUSTOMER WAS REPORTED THAT THE INSULIN PUMP DISPLAY DID NOT RETURN AFTER RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10153086 ↗
  • MalfunctionJun 10, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD KEYPAD THAT BUTTONS ARE EMBOSSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER STATED THAT NUMBERS ARE RAMPING ON THEIR OWN. CUSTOMER STATED THE SCROLL BAR WAS NOT MOVING WITH NO INPUT. CUSTOMER STATED THAT INSULIN PUMP WAS EXPOSED TO A CHANGE IN ALTITUDE. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10137263 ↗
  • MalfunctionJun 9, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A PARTIAL DISPLAY. CUSTOMER STATED THE ISSUE REMAINED UNRESOLVED RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10131924 ↗
  • MalfunctionJun 2, 2020Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE INSULIN PUMP ERROR ALARMS. CUSTOMER STATED THEY WERE UNABLE TO CLEAR THE ALARM. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10107592 ↗
  • MalfunctionMay 26, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR DID NOT LOCK IN PLACE INTO THE INSULIN PUMP. THE CUSTOMER ALSO STATED THAT THE RESERVOIR RETAINER RING WAS MISSING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10086513 ↗
  • MalfunctionMay 12, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP DISPLAY WAS CHANGING ITS COLOR AND THEY CANNOT SEE THINGS CLEARLY ON THE DISPLAY. THE SELF TEST WAS PERFORMED AND IT CONTINUES THE SAME. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10047795 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2981-2026 · Jul 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
  • Z-1742-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1261-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1265-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1266-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
  • Z-0193-2023 · Sep 20, 2022 · Open, Classified
    Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
  • Z-0855-2022 · Jan 27, 2022 · Open, Classified
    MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
  • Z-1771-2021 · Mar 15, 2021 · Open, Classified
    MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
  • Z-0516-2021 · Apr 30, 2020 · Open, Classified
    MiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
  • Z-0956-2020 · Nov 21, 2019 · Open, Classified
    Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
  • Z-2604-2017 · May 8, 2017 · Terminated
    MiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
  • Z-2605-2017 · May 8, 2017 · Terminated
    MiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.

Other Automated insulin dosing device system, single hormonal control devices

All Automated insulin dosing device system, single hormonal control devices →