MiniMed® 670G MMT-1780KL (UDI 00763000283513)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-1780KL 670G V4.13 JBL MG BC · Class 3
All models of MiniMed 670G (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic MiniMed 670G insulin pump delivers insulin for people with diabetes. Its 823 reports include 1 death and 279 injuries. Many describe battery problems, blank screens and pump errors, and some patients reported high or low blood sugar. 2 recalls name this device, both open: one about insulin over- or under-delivery when the pump is raised or lowered, and one about shorter battery life after the pump is dropped or bumped.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 823
- Deaths reported
- 1
- Injuries reported
- 279
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 823 adverse event reports that list this exact device (DI 00763000283513), received between Oct 3, 2019 and Sep 22, 2024 — classified as 1 death, 279 injuries, 543 malfunctions.
The most frequently reported specific device problem is “Power Problem” (180 reports).
2 FDA recall records name this device's identifier, 2 still open.
12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2026Open, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 31, 2024Open, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem213
- Power Problem180
- No Apparent Adverse Event122
- No Display/Image108
- Mechanical Problem89
- Patient Device Interaction Problem77
- Device Difficult to Program or Calibrate68
- Moisture or Humidity Problem67
Patient problems
- No Consequences Or Impact To Patient498
- Hyperglycemia204
- Hypoglycemia97
- Elevated Ketones/ Diabetic Ketoacidosis20
- No Clinical Signs, Symptoms or Conditions15
- Nausea9
- Insufficient Information8
- Vomiting8
Recent adverse event reports
The 25 most recent of 823 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2024Break; Moisture or Humidity Problem; Break; Moisture or Humidity Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THERE IS A CRACK AT THE BACK OF THE PUMP ON THE BATTERY COMPARTMENT AREA AND IT WAS REPORTED THAT CUSTOMER WENT SWIMMING NOT KNOWING THAT THERE IS A CRACK NOW THE PUMP WON¿T TURN ON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER DISCONTINUED USING THE PRODUCT AND PUMP WAS RETURNED FOR PRODUCT ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20276075 ↗MalfunctionJun 29, 2022Break; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC), BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14851026 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797279 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797101 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797037 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14796449 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14796008 ↗MalfunctionMay 18, 2022Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD PUMP ERROR. IT WAS REPORTED THAT THE CUSTOMER WAS ABLE TO CLEAR THE ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER DISCONTINUED THE USE OF DEVICE.
Patient problems: Hyperglycemia
MAUDE report 14427051 ↗MalfunctionMay 6, 2022Human-Device Interface Problem; Human-Device Interface Problem
IT WAS IDENTIFIED THAT THE CUSTOMER'S BASAL PROGRAMMING WAS UNKNOWN. THE CUSTOMER REPORTED THAT THEIR BLOOD GLUCOSE WAS RISING UP AND THEY FOUND THAT THE BASAL RATE WAS NOT SET UP. THE BLOOD GLUCOSE LEVEL AT THE TIME OF THE EVENT WAS 350 MG/DL. TROUBLESHOOTING WAS NOT PERFORMED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 14312937 ↗MalfunctionMay 6, 2022Human-Device Interface Problem; Human-Device Interface Problem
IT WAS IDENTIFIED THAT THE CUSTOMER'S BASAL PROGRAMMING WAS UNKNOWN. THE CUSTOMER REPORTED THAT THEY HAD BEEN HAVING HIGH BLOOD GLUCOSE AFTER CHANGING THE INFUSION SET. THE CUSTOMER'S BLOOD GLUCOSE AT THE TIME OF HIGH BLOOD GLUCOSE EVENT WAS 319 MG/DL. THE CUSTOMER DECLINED TROUBLESHOOTING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 14312482 ↗MalfunctionApr 20, 2022Human-Device Interface Problem; Human-Device Interface Problem
IT WAS IDENTIFIED THAT THE USER WAS NOT AWARE OR HAD FORGOTTEN HIS BASAL PROGRAMMING. THE CUSTOMER REPORTED THAT THEY EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER'S BLOOD GLUCOSE AT THE TIME OF CALL WAS 350MG/DL. CUSTOMER WAS USING THE INSULIN PUMP WITHIN 48 HOURS OF THE REPORTED HYPERGLYCEMIC EVENT. THE AUTO MODE FEATURE WAS NOT ACTIVE AT THE TIME OF THE HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER STARTED USING THE PUMP AND WAS ABLE TO DELIVER A BOLUS BEFORE EATING. TROUBLESHOOTING WAS PERFORMED AND IT WAS INDICATED THAT THE INSULIN PUMP WAS NOT PROGRAMMED.
Patient problems: Hyperglycemia
MAUDE report 14158024 ↗MalfunctionOct 28, 2021Battery Problem; Power Problem; Battery Problem; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD REPLACE BATTERY ALERT. CUSTOMER WAS CHANGING BATTERIES AND ALARM OCCURRED. THERE WAS SECOND OCCURRENCE OF REPLACE BATTERY NOW ALARM OR OFF NO POWER ALARM PRIOR LOW BATTERY ALERT. THE BATTERY CAP WAS NOT LOOSED OR DAMAGED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12710430 ↗MalfunctionOct 12, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12612829 ↗MalfunctionSep 20, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD DAMAGE ON THE RETAINER RING. CUSTOMER STATED THAT THE INSULIN PUMP WAS BROKEN BECAUSE OF THE ACCIDENT. IT WAS REPORTED THAT THE RESERVOIR ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12499346 ↗MalfunctionSep 16, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RETAINER RING AND PART OF PUMP BROKEN. THE RESERVOIR CAN LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12480535 ↗MalfunctionSep 15, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RETAINER RING DAMAGED. THE RESERVOIR CAN LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12474220 ↗MalfunctionSep 15, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD LOOSE RETAINER RING. THE CUSTOMER STATES THAT THE PUMP HAS NEITHER BEEN BUMPED NOR DROPPED. THE RESERVOIR CAN LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12473693 ↗MalfunctionSep 15, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RETAINER RING DAMAGED. THE RESERVOIR CAN LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12473497 ↗MalfunctionSep 15, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD DAMAGE ON THE RETAINER RING. IT WAS REPORTED THAT THE RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12472847 ↗MalfunctionSep 15, 2021Break; Mechanical Problem; Break; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD DAMAGE ON THE PLASTIC CLEAR RING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12472818 ↗MalfunctionSep 14, 2021Break; Patient Device Interaction Problem; Break; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RETAINER RING CRACKED. THE RESERVOIR CAN LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12466475 ↗MalfunctionJul 20, 2021Power Problem; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS HAVING BATTERY ISSUE. CUSTOMER WAS CALLING BACK AFTER COMPLETING PREVIOUS TROUBLESHOOTING. THE CUSTOMER STATED THEY HAVE BEEN GETTING A LOW BATTERY ALERT THAT LED TO REPLACE BATTERY NOW ALARM. CUSTOMER HAS BEEN CHANGING THE BATTERY MULTIPLE TIMES IN THE LAST 5 DAYS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12194730 ↗MalfunctionJun 8, 2021Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP RECEIVED MULTIPLE INSULIN PUMP ERRORS. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ABLE TO REWIND INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 11953664 ↗MalfunctionApr 9, 2021Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER WANTED TO MAKE SURE THAT THERE BASAL RATE WAS RUNNING IN THE INSULIN PUMP. CUSTOMER STATED THAT THEY NEEDS ASSISTANCE WITH THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 11641619 ↗MalfunctionMar 16, 2021Communication or Transmission Problem; Communication or Transmission Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A LOSS COMMUNICATION AND RECEIVED CANNOT FIND SENSOR SIGNAL ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11493019 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.
- Z-2981-2026 · Jul 13, 2026 · Open, ClassifiedMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Z-1742-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1261-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-1265-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1266-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-18
- Z-1771-2021 · Mar 15, 2021 · Open, ClassifiedMiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE
- Z-0516-2021 · Apr 30, 2020 · Open, ClassifiedMiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
- Z-0956-2020 · Nov 21, 2019 · Open, ClassifiedMedtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
- Z-2605-2017 · May 8, 2017 · TerminatedMiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.
- Z-2604-2017 · May 8, 2017 · TerminatedMiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
Other Automated insulin dosing device system, single hormonal control devices
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls78 deaths55,723 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 recalls51 deaths33,713 reports
- MiniMed®670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls39 deaths28,789 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths8,353 reports
- MiniMed® 670G MMT-1780KMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls7 deaths6,776 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls10 deaths4,325 reports
- MiniMed™ Reservoir MMT-332AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 deaths3,013 reports
- Guardian™ Sensor (3) MMT-7020AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control1,304 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page