Skip to content

Chocolate® CB1415025120OTW (UDI 00763000336707)Recalls & reported problems

Ev3, Inc · PTA CB1415025120 CHOCOLATE 014 V03 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000336707), received between Apr 29, 2022 and Apr 29, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    PHYSICIAN USED A CHOCOLATE PTA BALLOON DURING TREATMENT OF A CALCIFIED LESION IN THE PATIENT'S PROXIMAL POSTERIOR TIBIAL ARTERY.  MODERATE VESSEL TORTUOSITY AND CALCIFICATION ARE REPORTED. LESION EXHIBITED 60% STENOSIS. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED AS PER IFU WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT. IT IS REPORTED A BREAK/FRACTURE OCCURRED ON THE TIP OF THE DEVICE DURING ADVANCEMENT THROUGH THE GUIDE CATHETER. THE DETACHED TIP WAS REMOVED WITH THE CATHETER. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT BRAND DEVICE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14249592 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page