Chocolate™ CB1413540120OTWRecalls & reported problems

Medtronic, Inc. · PTA CB1413540120 CHOCOLATE 014 V05 · Class 2

All models of Chocolate (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000500115), received between Jul 28, 2025 and Nov 24, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Inflation Problem2
  • Detachment of Device or Device Component1
  • Device Difficult to Setup or Prepare1
  • Difficult to Remove1
  • Leak/Splash1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • Vascular Dissection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Device Difficult to Setup or Prepare; Difficult to Remove; Difficult to Advance; Positioning Problem; Device Difficult to Setup or Prepare; Difficult to Remove; Difficult to Advance; Positioning Problem

    IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A CHOCOLATE BALLOON TO TREAT A PLAQUE LESION IN THE SUPERFICIAL FEMORAL ARTERY. DEGREE OF TORTUOSITY MODERATE. DEGREE OF CALCIFICATION MODERATE. BALLOON WAS INFLATED WITH A BOSTON SCIENTIFIC PRESSURE PUMP. THERE WERE WIRE MOVEMENT ISSUES. DIFFICULTIES LOADING OR EXCHANGING GUIDEWIRE. POOR GUIDEWIRE MOVEMENT DURING PROCEDURE. DIFFICULTY REMOVING BALLOON PRIOR TO BALLOON INFLATION. RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE. THE SURGERY WAS COMPLETED AFTER REPLACING THE BALLOON WITH ONE OF THE SAME SIZE CHOCOLATE BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23631181 ↗
  • InjuryOct 23, 2025Detachment of Device or Device Component; Difficult to Advance; Detachment of Device or Device Component; Difficult to Advance

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE CHOCOLATE BALLOON, RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE THROUGH A SEVERELY CALCIFIED LESION (90% STENOSIS) LOCATED IN THE MID AND DISTAL SEGMENTS OF THE SUPERFICIAL FEMORAL ARTERY AND POPLITEAL ARTERY. THE ARTERY WAS MILDLY TORTUOUS. THE METAL WIRE OF THE CHOCOLATE BALLOON DETACHED AND COULD NOT BE REMOVED, A STENT WAS USED TO SECURE AND ATTACH THE RETAINED DEVICE COMPONENT WITHIN THE VESSEL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 23366653 ↗
  • InjurySep 5, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE TO TREAT SEVERE (90%) STENOSIS AND CALCIFICATION IN THE MID LEFT SUPERFICIAL FEMORAL ARTERY NEAR THE POPLITEAL FOSSA P1 SEGMENT, THERE WERE BALLOON INFLATION DIFFICULTIES AND THE CHOCOLATE BALLOON WOULD NOT OPEN. THE ARTERY WAS 4.5MM IN DIAMETER. THERE WAS MILD TORTUOSITY. ANGIOGRAPHY REVEALED SEVERE STENOSIS AND CALCIFICATION PRIOR TO INTERVENTION. PRE-DILATATION WAS PERFORMED WITH A 2.5X120 NON MEDTRONIC BALLOON. WHEN A 4X120 CHOCOLATE BALLOON WAS USED FOR FURTHER DILATATION, INFLATION WAS ATTEMPTED AT 12 ATM USING A PRESSURE PUMP, BUT THE BALLOON WAS DIFFICULT TO INFLATE AND DID NOT OPEN, SO IT WAS REMOVED. SUBSEQUENT DILATATION WITH A 4X120 NON MEDTRONIC BALLOON REVEALED A FLOW-LIMITING DISSECTION. A NON MEDTRONIC STENT WAS THEN INSERTED, RESTORING BLOOD FLOW AND COMPLETING THE PROCEDURE. IT IS UNCERTAIN IF THERE IS A LEAK AND IT IS UNCERTAIN WHAT THE CAUSE OF THE DISSECTION WAS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Vascular Dissection

    MAUDE report 22975622 ↗
  • MalfunctionJul 28, 2025Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE CHOCOLATE BALLOON, THE BALLOON WAS POORLY INFLATED AND COULD NOT FULLY DILATE THE LESION IN THE MID POPLITEAL ARTERY. A BALLOON BURST OR LEAK WAS OBSERVED DURING BALLOON INFLATION AT 4ATM. THE PROCEDURE WAS COMPLETED BY REPLACING THE DEVICE WITH ANOTHER OF THE SAME BRAND AND MODEL TO DILATE THE LESION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22630681 ↗