Skip to content

Fortrex™ A35HPV06040080 (UDI 00763000634490)Recalls & reported problems

Medtronic, Inc. · PTA A35HPV06040080 FORTREX V04 · Class 2

All models of Fortrex (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000634490), received between Feb 10, 2025 and Sep 2, 2026 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel7
  • Component Missing2
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50558288 ↗
  • MalfunctionSep 2, 2026Burst Container or Vessel

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50558030 ↗
  • MalfunctionNov 24, 2025Component Missing; Component Missing

    IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A FORTEX BALLOON TO TREAT A LESION IN THE ILIAC ARTERY - COMMON. THE DEVICE WAS MISSING A MARKER POINT. A NEW PRODUCT WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630499 ↗
  • MalfunctionNov 12, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE FOR STENOSIS DILATION USING A FORTREX BALLOON CATHETER, BALLOON LEAKAGE AND BURST WERE OBSERVED DURING THE SECOND INFLATION AT A PRESSURE OF 22 ATM. THE PATIENT WAS TREATED FOR A FIBROUS LESION WITH NO TORTUOSITY OR CALCIFICATION. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED, AND THE PROCEDURE WAS COMPLETED BY REPLACING THE DEVICE WITH ANOTHER BALLOON CATHETER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534518 ↗
  • MalfunctionAug 18, 2025Component Missing; Component Missing

    PHYSICIAN WAS ATTEMPTING TO USE FORTREX BALLOON TO TREAT PATIENT COMMON ILIAC ARTERY. IT WAS REPORTED THAT DEVICE WAS MISSING MARKER POINT. A NEW DEVICE WAS OPENED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22825914 ↗
  • MalfunctionJul 7, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE TO TREAT ARTERIOVENOUS FISTULA STENOSIS, THE FORTREX 06X040X80 BALLOON EXPERIENCED A LONGITUDINAL BALLOON BURST DURING INFLATION. A PRESSURE PUMP INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THE EVENT OCCURRED DURING BALLOON INFLATION AT A PRESSURE OF 24 ATM. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE PROCEDURE WAS COMPLETED USING A NEW BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22412476 ↗
  • MalfunctionJul 4, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE FORTREX 06X040X80 BALLOON, A PIN HOLE BURST WAS OBSERVED IN THE BALLOON DURING INFLATION AT 20 ATM. THE BALLOON LEAKED WAYER. THE PROCEDURE WAS BEING PERFORMED TO TREAT STENOSIS AT THE ANTERIOR WALL OF A PROSTHETIC VASCULAR GRAFT. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE.  ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PROCEDURE WAS COMPLETED USING A NEW BALLOON OF THE SAME SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22402368 ↗
  • MalfunctionJul 4, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PERIPHERAL PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING A FORTREX BALLOON. THE LESION WAS A PLAQUE LESION IN THE DISTAL REGION OF THE ARTERY. NO ISSUES NOTED WHEN THE DEVICE WAS REMOVED FROM PACKAGING. NO RESISTANCE DURING ADVANCEMENT. THE BALLOON WAS INFLATED USING A UNKNOWN PRESSURE PUMP. DURING BALLOON INFLATION A LONGITUDINAL BALLOON BURST OCCURRED AT AN INFLATION PRESSURE OF 24 ATM. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PROCEDURE WAS COMPLETED BY OPENING A NEW BOX OF BALLOONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22402361 ↗
  • InjuryMar 7, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING A FORTREX 06X040X80 BALLOON, A NON-MEDTRONIC INFLATION DEVICE WAS USED AND THE BALLOON BURST CIRCUMFERENTIALLY/RADIALLY DURING INFLATION AT A PRESSURE OF 24 ATMS. THE EVENT OCCURRED AT AN ARTERIOVENOUS FISTULA (AVF) SITE. A WOUND WAS OPENED TO ASSIST IN REMOVING THE BALLOON, AND ALL FRAGMENTS WERE SUCCESSFULLY RETRIEVED. THE PROCEDURE WAS COMPLETED USING A NEW HIGH-PRESSURE BALLOON FROM ANOTHER MANUFACTURER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21544815 ↗
  • MalfunctionFeb 10, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PROCEDURE THE ARTERIOSCLEROSIS WAS SEVERE, AND A PIN HOLE BURST OCCURRED WHEN THE FORTREX 06X040X80 BALLOON WAS INFLATED TO 12 BARS. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE PROCEDURE WAS COMPLETED BY REPLACING THE DEVICE WITH A NEW ONE AND CONTINUING THE OPERATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21342278 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page