The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000757700), received between Jan 22, 2025 and Mar 5, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem3
- Detachment of Device or Device Component2
- Burst Container or Vessel1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2026Deflation Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE NANOCROSS 014 DEVICE TO TREAT PERIPHERAL ARTERY PLAQUE IN THE RIGHT MID POSTERIOR TIBIAL ARTERY (70% STENOSIS, 250 MM LESION LENGTH, 3 MM ARTERY DIAMETER, MODERATE CALCIFICATION, MINIMAL TORTUOSITY), THERE WERE DELAYS IN SURGERY OF FIVE MINUTES DUE TO PRODUCT MALFUNCTION. DEFLATION DIFFICULTIES WERE ENCOUNTERED, AS THE BALLOON COULD NOT BE FULLY DEFLATED FOR REMOVAL. THE BALLOON DETACHED WITHIN THE SHEATH, AND CATHETER TOLERANCE WAS DEGRADED. THE SHEATH USED WAS A 6FR DESTINATION, AND THE GUIDEWIRE WAS A .014 GLADIUS. THE VESSEL WAS POST-DILATED WITH A NANO 3-2.5 DEVICE, AND INFLATION WAS PERFORMED WITH A PROGRAMMER 26 USING A 70 SALINE/30 CON FLUID. NEGATIVE PRESSURE AND PULLING WERE APPLIED TO DEFLATE THE BALLOON. THE DETACHED BALLOON COMPONENT WAS REMOVED WITHIN THE SHEATH, AND THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED BY REMOVING EVERYTHING WITH THE WIRE IN PLACE AND A NEW SHEATH TO FINISH. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24517131 ↗MalfunctionSep 2, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE NANOCROSS 014 BALLOON, A CHRONIC TOTAL OCCLUSION (100% STENOSIS) OF THE DISTAL TIBIAL/POPLIETAL TRUNK WITH MODERATE TORTUOSITY AND MODERATE CALCIFICATION WAS PRESENT. DURING BALLOON INFLATION, THE BALLOON BURST AND A CATHETER LEAK OCCURRED. THE DEVICE WAS PREPPED ACCORDING TO INSTRUCTIONS FOR USE. NO EMBOLIC PROTECTION WAS USED, AND NO RESISTANCE OR EXCESSIVE FORCE WAS ENCOUNTERED DURING DEVICE ADVANCEMENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22931727 ↗MalfunctionFeb 7, 2025Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY USING A NANOCROSS BALLOON, THERE WERE PROCEDURAL ISSUES RELATED TO THE DEFLATION OF THE BALLOON. THE BALLOON DID NOT DEFLATE AT THE TYPICAL RATE, REQUIRING MULTIPLE ATTEMPTS AND TAKING SEVERAL MINUTES TO DEFLATE. NEGATIVE PRESSURE WAS APPLIED TO AID IN THE DEFLATION PROCESS. DUE TO THE EXTENDED TIME REQUIRED TO WITHDRAW THE DEVICE, A NEW NANOCROSS BALLOON WAS USED. THE NEW BALLOON WAS INSPECTED, PREPPED, AND INFLATED WITHOUT FURTHER ISSUES. THE LESION LENGTH WAS SPECIFIED AS 150 MILLIMETERS, AND THE PROCEDURE WAS COMPLETED USING THE NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21332804 ↗InjuryJan 22, 2025Deflation Problem; Detachment of Device or Device Component; Deflation Problem; Detachment of Device or Device Component
PHYSICIAN ATTEMPTED TO USE NANOCROSS 014 BALLOON DILATATION DEVICE FOR PATIENT TREATMENT. A NON-MEDTRONIC SHEATH AND A NON-MEDTRONIC GUIDEWIRE AND A NON-MEDTRONIC INFLATION DEVICE WAS USED. THE DEVICE WAS PREPPED AS PER IFU HOWEVER ISSUES OCCURRED PREPPING THE DEVICE AND FLUSHING. IT WAS REPORTED THAT DEFLATION DIFFICULTIES ISSUES OCCURRED, AND DEVICE TOOK 5 MINS TO DEFLATE USING NEGATIVE PRESSURE WITH A 20ML SYRINGE WHICH CAUSED THE BALLOON TO DETACH IN THE VESSEL. THE DETACHED BALLOON WAS SNARED AND REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. NO FURTHER PATIENT INJURY WAS REPORTED FOR THIS EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21206319 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.
- Z-0904-2025 · Dec 10, 2024 · Open, ClassifiedMedtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page