Skip to content

NanoCross™ Elite AB14W040120150 (UDI 00763000757984)Recalls & reported problems

Medtronic, Inc. · PTA AB14W040120150 NANOCROSS 014 V06 · Class 2

All models of NanoCross Elite (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000757984), received between Feb 18, 2025 and Feb 18, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1
  • Difficult to Remove1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2025Deflation Problem; Inflation Problem; Difficult to Remove; Deflation Problem; Inflation Problem; Difficult to Remove

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY NANOCROSS 014 BALLOON, DEFLATION AND INFLATION DIFFICULTIES OCCURRED. THE BALLOON CATHETER EXPERIENCED DEFLATION DIFFICULTIES AND SLOWED INFLATION. THE BALLOON WAS NOT FULLY DEFLATED AND WAS REMOVED WITH DIFFICULTY. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE, WHICH WAS NOT A MEDTRONIC DEVICE. THE EVENT OCCURRED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, MID-LOCATION, WITH A LESION LENGTH OF 120MM AND AN ARTERY DIAMETER OF 5.1MM. THE SHEATH USED WAS A NON-MEDTRONIC 6F, AND THE INFLATION DEVICE WAS A NON-MEDTRONIC DEVICE WITH 50/50 CONTRAST FLUID. THE INSTRUCTIONS FOR USE WERE FOLLOWED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21397694 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.

  • Z-0904-2025 · Dec 10, 2024 · Open, Classified
    Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,

Other Catheter, angioplasty, peripheral, transluminal devices