Skip to content

NanoCross™ Elite AB14W050120150 (UDI 00763000758141)Recalls & reported problems

Medtronic, Inc. · PTA AB14W050120150 NANOCROSS 014 V06 · Class 2

All models of NanoCross Elite (30)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000758141), received between Jan 7, 2025 and Jan 7, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 10, 2024Class IIOpen, Classified

    There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.

    Z-0904-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Deflation Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 7, 2025Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING A NANOCROSS ELITE BALLOON, DEFLATION DIFFICULTIES WITH THE BALLOON WERE NOTED, THE DEVICE DID NOT FULLY DEFLATE FOR REMOVAL. A 5FR SHEATH AND A 0.014" NON MEDTRONIC GUIDEWIRE WERE USED. A NON MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST INFLATION FLUID. THE BALLOON WAS EVENTUALLY DEFLATED USING NEGATIVE PRESSURE AND BREAKING THROUGH THE SHEATH. THE PROCEDURE WAS ADDRESSING A LESION IN THE MID-LEFT POPLITEAL ARTERY WITH A LENGTH OF 120 MM, ASSOCIATED WITH PERIPHERAL ARTERY DISEASE AND PLAQUE. A NEW NON MEDTRONIC DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO HEALTH CONSEQUENCES OR IMPACT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21090128 ↗

FDA clearance and classification

Cleared through Special 510(k) K141118 on Jul 18, 2014 (decision: Substantially Equivalent). Submitted by Ev3, Inc. as “Nanocross Eite 0.014 Over-The-Wire Pta Balloon Dilatation Catheter”. The FDA review took 79 days. FDA record ↗

The FDA regulates product code LIT (catheter, angioplasty, peripheral, transluminal) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NanoCross™ Elite AB14W050120150

Has the NanoCross™ Elite AB14W050120150 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0904-2025 (Class II), began on Dec 10, 2024: There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NanoCross™ Elite AB14W050120150?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 7, 2025. That report lists device problem “Deflation Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NanoCross™ Elite AB14W050120150?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NanoCross™ Elite AB14W050120150 FDA approved?

It was cleared through Special 510(k) K141118 on Jul 18, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NanoCross™ Elite AB14W050120150, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NanoCross™ Elite, model AB14W050120150). Its record shows it as in commercial distribution.