NanoCross™ Elite AB14W050120150 (UDI 00763000758141)Recalls & reported problems
Medtronic, Inc. · PTA AB14W050120150 NANOCROSS 014 V06 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000758141), received between Jan 7, 2025 and Jan 7, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 10, 2024Open, Classified
There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.
Z-0904-2025 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Deflation Problem1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 7, 2025Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING A NANOCROSS ELITE BALLOON, DEFLATION DIFFICULTIES WITH THE BALLOON WERE NOTED, THE DEVICE DID NOT FULLY DEFLATE FOR REMOVAL. A 5FR SHEATH AND A 0.014" NON MEDTRONIC GUIDEWIRE WERE USED. A NON MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST INFLATION FLUID. THE BALLOON WAS EVENTUALLY DEFLATED USING NEGATIVE PRESSURE AND BREAKING THROUGH THE SHEATH. THE PROCEDURE WAS ADDRESSING A LESION IN THE MID-LEFT POPLITEAL ARTERY WITH A LENGTH OF 120 MM, ASSOCIATED WITH PERIPHERAL ARTERY DISEASE AND PLAQUE. A NEW NON MEDTRONIC DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO HEALTH CONSEQUENCES OR IMPACT REPORTED.
Patient problems: Insufficient Information
MAUDE report 21090128 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.