NanoCross™ Elite AB14W060150150 (UDI 00763000758325)Recalls & reported problems
Medtronic, Inc. · PTA AB14W060150150 NANOCROSS 014 V06 · Class 2
All models of NanoCross Elite (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000758325), received between Feb 19, 2025 and Dec 9, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 10, 2024Class IIOpen, Classified
There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.
Z-0904-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TREATMENT OF A FIBROUS LESION IN THE RIGHT PROXIMAL/MID SUPERFICIAL FEMORAL ARTERY WITH A NANOCROSS ELITE PTA BALLOON, THERE WAS A ONE-HOUR DELAY IN SURGERY DUE TO A DETACHED BALLOON AT THE TARGET LESION. ARTERY DIAMETER REPORTED AS 6MM. A 6FR NON-MEDTRONIC SHEATH WAS USED. A NON-MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST FLUID FOR BALLOON INFLATION. THE DEVICE WAS PASSED THROUGH A PREVIOUSLY DEPLOYED 6MM PROTEGE EVERFLEX STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE VESSEL WAS POST-DILATED. THE DETACHED BALLOON WAS SNARED AND REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW MEDTRONIC DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23755875 ↗MalfunctionFeb 19, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT A CALCIFIED LESION IN THE MID RIGHT POPLITEAL ARTERY USING A NANOCROSS ELITE, A PINHOLD BURST OCCURRED AT 10 ATMS DURING THE FIRST INFLATION. THE VESSEL HAD NOT BEEN PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. 50/50 CONTRAST FLUID WAS USED FOR INFLATION. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO BALLOON DETACHMENT OCCURRED. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21411439 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page