EverCross™ AB35W07060135 (UDI 00763000766757)Recalls & reported problems
Medtronic, Inc. · PTA AB35W07060135 EVERCROSS 035 V10 · Class 2
All models of EverCross (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000766757), received between Sep 2, 2025 and Sep 2, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 2, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE EVERCROSS 035 BALLOON, A PATIENT WITH A SEVERELY CALCIFIED LESION IN THE COMMON ILIAC ARTERY (RIGHT AND LEFT, PROXIMAL AND MID SEGMENTS) EXPERIENCED BALLOON BURST ON FIRST INFLATION AT OR BELOW THE RATED BURST PRESSURE. A 6X11CM NON-MEDTRONIC SHEATH WAS USED. A NON-MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST. THE BALLOON BURST OCCURRED WITH TWO SEPARATE EVERCROSS 035 DEVICES. THERE WAS NO INJURY TO THE PATIENT ASSOCIATED WITH THE BALLOON BURST, AND THE DEVICES WERE SAFELY REMOVED FROM THE PATIENT AFTER EACH EVENT. THE PROCEDURE WAS COMPLETED USING A NEW 7X60 EVERCROSS DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22932051 ↗MalfunctionSep 2, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE EVERCROSS 035 BALLOON, A PATIENT WITH A SEVERELY CALCIFIED LESION IN THE COMMON ILIAC ARTERY (RIGHT AND LEFT, PROXIMAL AND MID SEGMENTS) EXPERIENCED BALLOON BURST ON FIRST INFLATION AT OR BELOW THE RATED BURST PRESSURE. A 6X11CM NON MEDTRONIC SHEATH WAS USED. A NON MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST. THE BALLOON BURST OCCURRED WITH TWO SEPARATE EVERCROSS 035 DEVICES. THERE WAS NO INJURY TO THE PATIENT ASSOCIATED WITH THE BALLOON BURST, AND THE DEVICES WERE SAFELY REMOVED FROM THE PATIENT AFTER EACH EVENT. THE PROCEDURE WAS COMPLETED USING A NEW 7X60 EVERCROSS DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22931742 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page