Skip to content

EverCross™ AB35W08080135 (UDI 00763000766986)Recalls & reported problems

Medtronic, Inc. · PTA AB35W08080135 EVERCROSS 035 V10 · Class 2

All models of EverCross (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000766986), received between Feb 5, 2025 and Feb 5, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Device Damaged by Another Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 5, 2025Burst Container or Vessel; Device Damaged by Another Device; Burst Container or Vessel; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE EVERCROSS 035 BALLOONS, THE BALLOONS BURST DURING THE PROCEDURE AT A PRESSURE OF 8 ATMS ON THE FIRST INFLATION. THE PROCEDURE INVOLVED THE RIGHT PROXIMAL ILIAC ARTERY, WHICH WAS CALCIFIED WITH 80% STENOSIS AND A DIAMETER OF 8 MILLIMETERS. THE DEGREE OF TORTUOSITY WAS MINIMAL. A 6FR NON-MEDTRONIC SHEATH AND A 0.035" NON-MEDTRONIC GUIDEWIRE WERE USED. THE DEVICES WERE PASSED THROUGH A PREVIOUSLY DEPLOYED STENT WITHOUT ENCOUNTERING RESISTANCE. THE BALLOONS WERE INFLATED USING AN INFLATION DEVICE WITH A 50/50 CONTRAST FLUID. THE DEVICES WERE SAFELY REMOVED FROM THE PATIENT, AND THERE WERE NO INJURIES OR INTERVENTIONS ASSOCIATED WITH THE EVENT. THE PHYSICIAN BELIEVES THAT BOTH BALLOONS LIKELY GOT CAUGHT ON THE STRUT OF THE PREVIOUSLY IMPLANTED STENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21313209 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →