Ultraverse® RX U41502H4RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 2.5 mm x 40 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741056475), received between Oct 27, 2020 and Jul 9, 2024 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Detachment of Device or Device Component3
  • Entrapment of Device2
  • Material Deformation2
  • Device-Device Incompatibility1
  • Difficult to Open or Remove Packaging Material1
  • Difficult to Remove1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2024Difficult to Open or Remove Packaging Material; Material Deformation; Difficult to Open or Remove Packaging Material; Material Deformation

    IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE TIP OF THE STYLET WAS ATTACHED TO THE PTA BALLOON AND IT ALLEGEDLY WOULD NOT COME OFF. IT WAS FURTHER REPORTED THAT THE TIP OF THE STYLET WAS ALLEGEDLY CRUSHED AND COULD NOT BE USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19699516 ↗
  • MalfunctionMar 25, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE SHAFT OF THE CATHETER ALLEGEDLY KINKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18970625 ↗
  • MalfunctionFeb 20, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED BEFORE NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18738973 ↗
  • InjuryOct 26, 2023Entrapment of Device; Failure to Advance; Detachment of Device or Device Component; Entrapment of Device; Failure to Advance; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE POPLITEAL ARTERY VIA A CONTRALATERAL APPROACH, THE CATHETER APPARENTLY GOT STUCK WITH THE STENT PLACED IN THE SUPERFICIAL FEMORAL ARTERY AND DETACHED BEFORE REACHING THE LESION. IT WAS FURTHER REPORTED THAT THE DETACHED SEGMENT WAS UNABLE TO BE REMOVED USING SNARE. REPORTEDLY, THE DETACHED SEGMENT WAS REMOVED BY SURGICALLY. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18009140 ↗
  • MalfunctionMar 30, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A HIGHLY CALCIFIED TARGET LESION IN THE ANTERIOR TIBIAL ARTERY VIA AN IPSILATERAL ANTEGRADE APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED UNDER THE RATED BURST PRESSURE. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16653014 ↗
  • MalfunctionFeb 15, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16376740 ↗
  • MalfunctionJan 19, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A HIGHLY CALCIFIED TARGET LESION IN THE ANTERIOR TIBIAL ARTERY VIA AN IPSILATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 6 ATM. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16200731 ↗
  • InjuryMay 15, 2022Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component; Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION IN THE RIGHT ANTERIOR TIBIAL ARTERY VIA CONTRALATERAL APPROACH, THE PTA BALLOON WAS ALLEGEDLY TRAPPED BY THE CALCIFICATION OF THE LESION AND UNABLE TO COME OFF. IT WAS FURTHER REPORTED THAT THE SHAFT WAS DETACHED FROM THE TIP AND FRAGMENTS REMAINED IN THE PATIENTS BODY. REPORTEDLY, THE RIGHT COMMON FEMORAL ARTERY WAS INCISED AND THE BALLOON SHAFT WAS TAKEN OUT. THE CURRENT PATIENT STATUS IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 14403833 ↗
  • MalfunctionApr 15, 2021Material Discolored; Leak/Splash; Detachment of Device or Device Component; Device-Device Incompatibility; Material Discolored; Leak/Splash; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY LEAKED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS DISCOLORED AND UNABLE TO DELIVER OVER GUIDEWIRE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11669876 ↗
  • MalfunctionOct 27, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U41502H4RX PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE AGE AND WEIGHT OF MALE PATIENT WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10744034 ↗

Other Catheter, angioplasty, peripheral, transluminal devices