Skip to content

Ultraverse® RX U42002H20RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 2.5 mm x 200 mm, 200 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741056666), received between May 24, 2022 and Jun 24, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device-Device Incompatibility1
  • Failure to Advance1
  • Material Deformation1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2025Break; Break

    ON (B)(6) 2025, AN 88-YEAR-OLD MALE PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE RX PTA BALLOON CATHETER. PRIOR TO THE PROCEDURE, THE CATHETER SHAFT BROKE OUTSIDE THE BODY. IT WAS ABLE TO ABLE TO WITHDRAWAL THE TIP WITHOUT INCIDENCE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22322164 ↗
  • MalfunctionMay 24, 2022Failure to Advance; Device-Device Incompatibility; Material Deformation; Material Split, Cut or Torn; Failure to Advance; Device-Device Incompatibility; Material Deformation; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA CONTRALATERAL FEMORAL ARTERY, THE WIRE WAS ALLEGEDLY FLOATED AROUND THE LOOP. IT WAS FURTHER REPORTED THAT THE BALLOON ALLEGEDLY WOULD NOT TRACK AROUND THE HEEL. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14487024 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page