Ultraverse® RX U41502H25RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 2.5 mm x 250 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741056680), received between Jan 15, 2020 and Sep 16, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Deflation Problem2
  • Detachment of Device or Device Component2
  • Failure to Advance2
  • Material Deformation2
  • Break1
  • Device-Device Incompatibility1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025Break; Deflation Problem; Difficult to Insert; Difficult to Remove; Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility; Break; Deflation Problem; Difficult to Insert; Difficult to Remove; Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE RX PTA BALLOON CATHETER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PATIENT EXPERIENCED SOME RESISTANCE AND DISCOMFORT DURING INSERTION. ADDITIONALLY, THE CATHETER COULD NOT BE DEFLATED AFTER USE AND EVEN APPLYING NEGATIVE PRESSURE WITH AN INDEFLATOR SYRINGE FAILED TO REMOVE IT. AN ATTEMPT WAS MADE TO REMOVE THE ENTIRE SHEATH, BUT THE SHAFT BROKE OFF. THE WIRE REMAINED IN THE LUMEN, SO IT WAS RETRIEVED USING A GUIDING CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23060662 ↗
  • MalfunctionMay 3, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN AN ANTERIOR TIBIAL ARTERY LESION WITH CALCIFICATION IN THE IPSILATERAL APPROACH, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT NOMINAL 6 ATM PRESSURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19236905 ↗
  • MalfunctionMay 18, 2022Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE DAMAGE WAS FOUND AT THE PROXIMAL PART AFTER REMOVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14433262 ↗
  • MalfunctionMay 25, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT NOMINAL PRESSURE. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11875183 ↗
  • MalfunctionMar 5, 2021Deflation Problem; Inflation Problem; Material Deformation; Deflation Problem; Inflation Problem; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY FAILED TO INFLATE. IT WAS FURTHER REPORTED THAT THE BALLOON SUCCESSFULLY INFLATED OUTSIDE THE PATIENT'S BODY, BUT ALLEGEDLY FAILED TO DEFLATE. REPORTEDLY, ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11419502 ↗
  • MalfunctionJan 15, 2020Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U41502H25RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL DEFORMATION AND FAILURE TO ADVANCE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT WAS A (B)(6) YEAR OLD MALE; WEIGHT WAS NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9591799 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →