Ultraverse® RX U415034RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 3 mm x 40 mm, 150 cm Catheter · Class 2

Quick summary

This Bard Ultraverse RX balloon catheter is used to open narrowed peripheral arteries during angioplasty. Most of the 26 reports describe the balloon bursting during the procedure, sometimes below normal pressure, plus a few cases of the balloon deforming, breaking off at the hub, or being hard to remove. No deaths or injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00801741056802), received between Jul 22, 2019 and Feb 17, 2026 — classified as 26 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture21
  • Material Deformation3
  • Detachment of Device or Device Component2
  • Difficult to Remove2
  • Failure to Advance2
  • Break1
  • Device-Device Incompatibility1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • No Consequences Or Impact To Patient8

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Material Rupture

    ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE RX BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24363360 ↗
  • MalfunctionJan 29, 2026Material Rupture

    ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE BY USING THE ULTRAVERSE RX BALLOON CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24206853 ↗
  • MalfunctionOct 16, 2025Material Rupture; Material Rupture

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE ULTRAVERSE RX PTA BALLOON CATHETER VIA IPSILATERAL APPROACH. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23307620 ↗
  • MalfunctionMar 11, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON WAS ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21570071 ↗
  • MalfunctionJan 11, 2025Break; Difficult to Remove; Detachment of Device or Device Component; Break; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA IPSILATERAL ANTEGRADE APPROACH, THE PTA BALLOON WAS ALLEGEDLY BROKEN OFF AT THE HUB PART AT THE BASE OF THE SHAFT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21132753 ↗
  • MalfunctionJan 5, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE MODERATELY CALCIFIED STENOTIC LESION, THE PTA BALLOON ALLEGEDLY RUPTURED BELOW NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21076060 ↗
  • MalfunctionJun 12, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL ARTERY LESION VIA CONTRALATERAL APPROACH, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19517386 ↗
  • MalfunctionFeb 13, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED DURING DILATION AT 6 BAR. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18694523 ↗
  • MalfunctionNov 27, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING A ANGIOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY RUPTURED AT 8 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18210702 ↗
  • MalfunctionSep 8, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED AND OCCLUDED TARGET LESION BELOW THE KNEE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17708906 ↗
  • MalfunctionFeb 2, 2023Leak/Splash; Material Rupture; Material Split, Cut or Torn; Leak/Splash; Material Rupture; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION OF CHRONIC TOTAL OCCLUSION IN THE SUPERFICIAL FEMORAL ARTERY VIA A RIGHT CONTRALATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 2 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16281955 ↗
  • MalfunctionMay 31, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION IN THE POPLITEAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED AROUND RATED BURST PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14547608 ↗
  • MalfunctionApr 6, 2022Difficult to Remove; Detachment of Device or Device Component; Device-Device Incompatibility; Difficult to Remove; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE CEPHALIC VEIN, THE PTA BALLOON ALLEGEDLY GOT STUCK IN THE GUIDEWIRE AND WAS DIFFICULT TO REMOVE WHEN THE PHYSICIAN ADJUSTED THE POSITION OF THE BALLOON. IT WAS FURTHER REPORTED THAT THE BALLOON AND GUIDEWIRE WERE REMOVED TOGETHER AS ONE UNIT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14020646 ↗
  • MalfunctionDec 21, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A HIGHLY CALCIFIED TARGET LESION IN THE LEFT ANTERIOR TIBIAL ARTERY VIA AN IPSILATERAL ANTEGRADE APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13051811 ↗
  • MalfunctionMay 11, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED AT 10 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11805750 ↗
  • MalfunctionApr 7, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT SUPERFICIAL FEMORAL ARTERY VIA THE IPSILATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 4 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11626249 ↗
  • MalfunctionMar 8, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF HIGHLY CALCIFIED LEGION, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT ANOTHER PTA BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431132 ↗
  • MalfunctionDec 11, 2020Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CATHETER SHAFT WAS ALLEGEDLY KINKED AND UNABLE TO CROSS LESION. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10990460 ↗
  • MalfunctionOct 27, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO CONSEQUENCES. THE MALE PATIENT'S AGE AND WEIGHT WAS NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10738948 ↗
  • MalfunctionOct 27, 2020Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THE MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL DEFORMATION, AND FAILURE TO ADVANCE. THE REPORT WAS RECEIVED FROM A SINGLE SOURCE. THIS MALFUNCTION INVOLVED PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT'S AGE, GENDER, AND WEIGHT WAS NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10738913 ↗
  • MalfunctionJul 22, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON RUPTURE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. THESE MALFUNCTIONS INVOLVED PATIENTS WITH NO CONSEQUENCES. TWO PATIENTS RANGED FROM 76-83 YEARS OF AGE AND 63-70 KGS. OF THE REPORTED PATIENTS, TWO WAS MALE AND ONE WERE FEMALE. ALL REMAINING PATIENT INFORMATION WAS NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10311009 ↗
  • MalfunctionApr 15, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. BOTH EVENTS DID INVOLVE PATIENTS WITH NO REPORTED PATIENT INJURY. ONE PATIENT IS (B)(6) OLD, (B)(6) , AND FEMALE; HOWEVER, THE SECOND PATIENTS¿ WEIGHT, GENDER AND AGE WERE NOT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9963690 ↗
  • MalfunctionJan 15, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX ULTRAVERSE RX PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES. OUT OF THE TWO REPORTED MALFUNCTIONS, BOTH PATIENTS WERE INVOLVED WITH NO PATIENT CONSEQUENCE OR IMPACT. ONE PATIENT WAS A MALE, NO AGE AND WEIGHT INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9594167 ↗
  • MalfunctionOct 22, 2019Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED AND THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT IS FEMALE AND AGE AND WEIGHT WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9219897 ↗
  • MalfunctionOct 21, 2019Material Deformation; Material Deformation

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U415034RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL DEFORMATION. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION DID INVOLVE A PATIENT WITH NO PATIENT INJURY. AGE, WEIGHT, AND GENDER WERE NOT PROVIDED FOR THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9216282 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →