Skip to content

Ultraverse® RX U4150330RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 3 mm x 300 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741057045), received between Nov 13, 2018 and Apr 16, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Break1
  • Detachment of Device or Device Component1
  • Material Deformation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • Rupture1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN A SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY, THE ULTRAVERSE RX PTA BALLOON WAS ALLEGEDLY RUPTURED AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21847583 ↗
  • MalfunctionJul 23, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19811973 ↗
  • MalfunctionFeb 27, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN A HIGHLY CALCIFIED LESION VIA IPSILATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18786918 ↗
  • MalfunctionJul 6, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION, THE PTA BALLOON ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14931665 ↗
  • InjuryJul 1, 2021Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE SHAFT OF THE CATHETER WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE VESSEL ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THE PATIENT CURRENT STATUS WAS UNKNOWN.

    Patient problems: Rupture

    MAUDE report 12098298 ↗
  • MalfunctionJul 21, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL U4150330RX PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THE INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE TWO MALFUNCTIONS, BOTH INVOLVED PATIENTS WITH NO PATIENT CONSEQUENCES. THE (B)(6) YEAR OLD MALE PATIENT WEIGHT WAS NOT PROVIDED. THE AGE, WEIGHT, AND GENDER WERE NOT PROVIDED FOR ANOTHER PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10305959 ↗
  • MalfunctionNov 13, 2018Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT SFA AND POPLITEAL ARTERY, THE PTA BALLOON CATHETER SHAFT ALLEGEDLY BROKE AFTER THE FIRST INFLATION ATTEMPT. THERE WAS NO REPORTED RETRACTION ISSUE THROUGH THE 6FR SHEATH. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8066335 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page