Skip to content

Ultraverse® RX U4150425RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 4 mm x 250 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741057670), received between Apr 15, 2020 and Jun 28, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation3
  • Material Rupture3
  • Failure to Advance2
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA ANTEGRADE APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED DURING THE FIRST DILATION 10 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19631960 ↗
  • MalfunctionMar 14, 2023Failure to Advance; Device-Device Incompatibility; Material Deformation; Failure to Advance; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA AN IPSILATERAL ANTEGRADE APPROACH, THERE WAS STRONG RESISTANCE FELT AND THE DEVICE ALLEGEDLY FAILED TO BE ADVANCED OVER THE GUIDEWIRE. IT WAS FURTHER REPORTED THAT THE SHAFT OF THE CATHETER WAS ALLEGEDLY KINKED AND WIRE LUMEN CRUSHED. THE BALLOON AND THE GUIDEWIRE WERE REMOVED TOGETHER AS ONE UNIT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16535323 ↗
  • MalfunctionJul 22, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING PREPARATION FOR AN ANGIOPLASTY PROCEDURE, THE CATHETER SHAFT WAS ALLEGEDLY KINKED . IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12208191 ↗
  • MalfunctionJan 22, 2021Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CATHETER SHAFT WAS ALLEGEDLY KINKED AND THE PTA BALLOON WAS ALLEGEDLY UNABLE TO CROSS THE LESION. IT WAS FURTHER REPORTED THAT ANOTHER PTA BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11213839 ↗
  • MalfunctionNov 3, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF LONG CHRONIC TOTAL OCCLUSION OF CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY VIA IPSILATERAL ANTE GRADE APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 6 ATM. IT WAS FURTHER REPORTED THAT ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10782975 ↗
  • MalfunctionApr 15, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U4150425RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. THE MALFUNCTIONS EACH INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. ONE PATIENT WAS MALE; ALL OTHER PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9963414 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page