Ultraverse® RX U4150515RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 5 mm x 150 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00801741057946), received between Dec 8, 2016 and Jul 11, 2025 — classified as 1 injury, 19 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture18
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Device-Device Incompatibility1
  • Difficult to Remove1
  • Inflation Problem1
  • Kinked1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient3
  • Intimal Dissection1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 11, 2025Material Rupture; Material Rupture

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE RX PTA BALLOON CATHETER TO TREAT CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH. DURING THE PROCEDURE, THE BALLOON RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22478333 ↗
  • MalfunctionJul 2, 2025Material Rupture; Material Rupture

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE RX PTA BALLOON CATHETER. DURING THE PROCEDURE THE BALLOON WAS ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22383752 ↗
  • MalfunctionApr 1, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE PTA BALLOON WAS ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21737340 ↗
  • MalfunctionFeb 27, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE IPSILATERAL ANTEROPOSTERIOR TO THE SUPERFICIAL FEMORAL ARTERY CALCIFIED LESION, THE PTA BALLOON ALLEGEDLY DILATED AT NOMINAL PRESSURE AND RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21477253 ↗
  • MalfunctionJul 10, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19709461 ↗
  • MalfunctionJun 10, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY HAD A PINHOLE RUPTURE AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19496046 ↗
  • MalfunctionMay 29, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19413844 ↗
  • MalfunctionApr 3, 2024Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY FAILED TO INFLATE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19032254 ↗
  • MalfunctionMar 5, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 8 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18843054 ↗
  • MalfunctionFeb 27, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA THE IPSILATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AT 7 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18786084 ↗
  • MalfunctionSep 4, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA AN IPSILATERAL ANTEGRADE APPROACH, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT FOUR ATMOSPHERIC PRESSURES. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17676006 ↗
  • MalfunctionJan 10, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16130249 ↗
  • MalfunctionAug 13, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF CHRONIC TOTAL OCCLUSION IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15222461 ↗
  • MalfunctionDec 12, 2021Difficult to Remove; Material Separation; Device-Device Incompatibility; Difficult to Remove; Material Separation; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE GUIDEWIRE ALLEGEDLY GOT STUCK WITH THE RECANALIZATION CATHETER. IT WAS FURTHER REPORTED THAT THE CATHETER AND GUIDEWIRE WERE REMOVED TOGETHER AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12981971 ↗
  • MalfunctionJan 19, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF CALCIFIED LESION IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11192392 ↗
  • MalfunctionDec 28, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF THE LESION OF CHRONIC TOTAL OCCLUSION IN THE SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11075679 ↗
  • MalfunctionJul 22, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U4150515RX DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON RUPTURE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE TWO MALFUNCTIONS, TWO PATIENTS WERE INVOLVED WITH NO PATIENT CONSEQUENCES OR IMPACT. THE TWO PATIENTS INVOLVED WERE MALE, NO DETAILS WERE PROVIDED ON ONE OF THE PATIENTS AND THE OTHER PATIENT WAS REPORTED TO BE (B)(6) YEARS OLD AND WEIGHED (B)(6) KGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10311217 ↗
  • MalfunctionJan 15, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U4150515RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCE. THE PATIENT¿S AGE, WEIGHT, AND SEX WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9593580 ↗
  • MalfunctionDec 5, 2017Break; Kinked; Material Rupture; Break; Kinked; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA, THE PTA BALLOON ALLEGEDLY RUPTURED AT 4 ATM. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7084860 ↗
  • InjuryDec 8, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE BETWEEN THE RIGHT SFA AND THE PROXIMAL POPLITEAL ARTERY, VESSEL DISSECTION WAS ALLEGEDLY REVEALED POST INFLATION. REPORTEDLY, A STENT WAS SUCCESSFULLY PLACED AT THE LESION SITE AND POST-DILATATION WAS PERFORMED WITH A PTA BALLOON CATHETER. THE VESSEL WAS PATENT WITH GOOD BLOOD FLOW AT THE CONCLUSION OF THE PROCEDURE. THERE WAS NO KNOWN REPORTED IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: Intimal Dissection

    MAUDE report 6156708 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →