Ultraverse® RX U4150525RXRecalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® RX PTA Dilatation Catheter 5 mm x 250 mm, 150 cm Catheter · Class 2

All models of Ultraverse RX (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741058004), received between Apr 4, 2017 and Jul 16, 2024 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture7
  • Material Deformation2
  • Break1
  • Detachment of Device or Device Component1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA CONTRALATERAL SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19748728 ↗
  • MalfunctionJun 25, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE CALCIFIED SUPERFICIAL FEMORAL ARTERY LESION VIA THE SAME SIDE APPROACH, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 8 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19605141 ↗
  • MalfunctionAug 8, 2022Failure to Advance; Material Deformation; Failure to Advance; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CATHETER SHAFT OF THE PTA BALLOON ALLEGEDLY KINKED AS THE PHYSICIAN APPLIED FORCE TO ADVANCE THE BALLOON. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15181644 ↗
  • MalfunctionApr 21, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CATHETER SHAFT OF THE PTA BALLOON ALLEGEDLY KINKED AS THE PHYSICIAN APPLIED STRONG FORCE AT THE LESION. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14166969 ↗
  • MalfunctionApr 21, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF CALCIFIED LESION, THE PTA BALLOON ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11698687 ↗
  • MalfunctionDec 15, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT ANOTHER PTA BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11010009 ↗
  • MalfunctionJul 22, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES THREE MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U4150525RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON RUPTURE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. ALL THREE MALFUNCTIONS INVOLVED A PATIENT WITH NO REPORTED PATIENT CONSEQUENCES. THE PATIENT'S AGE, WEIGHT AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10311424 ↗
  • MalfunctionApr 17, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U4150525RX PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO CONSEQUENCES. THE PATIENT'S AGE, WEIGHT AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9975246 ↗
  • MalfunctionApr 7, 2017Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA, THE PTA BALLOON ALLEGEDLY RUPTURED AT 3 ATM ON THE FIRST INFLATION ATTEMPT. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6474250 ↗
  • MalfunctionApr 4, 2017Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED DURING AN ANGIOPLASTY PROCEDURE USING A CONTRALATERAL APPROACH WITH INCREASING DILATING PRESSURE, THE PTA BALLOON CATHETER ALLEGEDLY BROKE WHEN IT WAS ADVANCED ACROSS THE VESSEL WITH FORCE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6457371 ↗

Other Catheter, angioplasty, peripheral, transluminal devices