Bard Peripheral Vascular, Inc. · Conquest® 40 PTA Dilatation Catheter 10 mm x 80 mm, 75 cm catheter · Class 2

All models of Conquest 40 (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741060649), received between Jan 22, 2018 and Sep 28, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation4
  • Material Rupture4
  • Difficult to Remove3
  • Inflation Problem2
  • Break1
  • Device Contamination with Chemical or Other Material1
  • Entrapment of Device1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Material Rupture

    On (b)(6) 2026, a patient underwent an angioplasty procedure using a Conquest 40 PTA balloon. During the procedure, the balloon was successfully inflated to dilate a stent. After deflation and withdrawal for repositioning to another treatment site, the physician attempted to reinflate the balloon but observed that it would not maintain pressure. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911163 ↗
  • MalfunctionOct 8, 2025Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST 40 PTA BALLOON. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY INVERTED ALONG THE PROXIMAL SEGMENT. IT WAS FURTHER REPORTED THAT IT COULD NOT BE REMOVED FROM THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23244009 ↗
  • MalfunctionJul 8, 2025Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST 40 PTA BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON MUSHROOMED ON THE DISTAL END AND WAS DIFFICULT TO REMOVE. IT WAS FURTHER REPORTED THAT MANEUVERED THE BALLOON SEVERAL WAYS IN ORDER TO EVENTUALLY REMOVE IT AND RESULTED IN CASE BEING DELAYED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22434867 ↗
  • MalfunctionJun 10, 2025Difficult to Remove; Difficult to Remove

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST 40 PTA BALLOON CATHETER. DURING THE PROCEDURE, WHEN PULLING BACK THE BALLOON RESISTANCE WAS FELT SO IT WAS PULLED WITH MORE FORCE AND BALLOON WAS UNSHEATHED AND STUCK AT THE END OF THE G-TUBE. IT WAS FURTHER REPORTED THAT G-TUBE AND BALLOON REMOVED FROM PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22174889 ↗
  • MalfunctionJul 9, 2021Inflation Problem; Material Rupture; Material Deformation; Inflation Problem; Material Rupture; Material Deformation

    IT WAS REPORTED DURING A TREATMENT OF STENOSIS IN THE SUPERIOR VENA CAVA JUST BEYOND THE CEPHALIC ARCH, THE BALLOON ALLEGEDLY RUPTURED AT 20 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12142599 ↗
  • MalfunctionJul 9, 2021Material Rupture; Material Deformation; Material Rupture; Material Deformation

    IT WAS REPORTED THAT DURING A TREATMENT OF STENOSIS IN THE SUPERIOR VENA CAVA JUST BEYOND THE CEPHALIC ARCH, THE BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE BALLOON ALLEGEDLY DEFORMED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12142595 ↗
  • MalfunctionJul 24, 2019Entrapment of Device; Inflation Problem; Device Contamination with Chemical or Other Material; Entrapment of Device; Inflation Problem; Device Contamination with Chemical or Other Material

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF75108 PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED ENTRAPMENT OF THE DEVICE, INFLATION ISSUES, RETRACTION PROBLEMS, MATERIAL RUPTURE, FAILURE TO ADVANCE AND FOREIGN MATERIAL. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCES. THE PATIENT WAS A 71-YEAR-OLD FEMALE WHO WEIGHED 166 POUNDS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8823498 ↗
  • MalfunctionJul 16, 2019Break; Material Puncture/Hole; Retraction Problem; Break; Material Puncture/Hole; Retraction Problem

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF75108 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A BREAK, MATERIAL PUNCTURE, AND A RETRACTION PROBLEM. THIS INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCE. THE PATIENT¿S AGE, WEIGHT, AND SEX WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8796346 ↗
  • MalfunctionJan 22, 2018Material Frayed; Material Rupture; Material Frayed; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 20 ATM. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7211383 ↗

Other Catheter, angioplasty, peripheral, transluminal devices