Skip to content

Atlas® Gold ATG120244Recalls & reported problems

Bard Peripheral Vascular, Inc. · Atlas® Gold PTA Dilatation Catheter 24 mm x 40 mm, 120 cm catheter · Class 2

All models of Atlas Gold (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741061028), received between Oct 18, 2017 and Oct 22, 2024 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Detachment of Device or Device Component3
  • Improper or Incorrect Procedure or Method2
  • Retraction Problem2
  • Break1
  • Difficult to Remove1
  • Leak/Splash1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Pericardial Effusion1
  • Vascular Dissection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 22, 2024Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING ANGIOPLASTY PROCEDURE, THE PTA BALLOON CAPSULE ALLEGEDLY FELL OFF DURING THE OPERATION AND COULD NOT BE REMOVED. IT WAS FURTHER REPORTED THAT THE CAPSULE CANNOT BE CAPTURED THROUGH INTERVENTIONAL TECHNIQUES. REPORTEDLY, THE PATIENT WAS IN THE CARDIAC SURGERY OF THE OPEN THORACIC SURGERY TO TAKE OUT THE BALLOON. THE CURRENT STATUS OF THE PATIENT IS STABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20506763 ↗
  • MalfunctionDec 11, 2023Peeled/Delaminated; Material Rupture; Peeled/Delaminated; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 10 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18300735 ↗
  • MalfunctionNov 14, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE FOR AN AORTIC VALVE REPLACEMENT, VIA THE FEMORAL ACCESS, THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE PUNCTURED WITH A HOLE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18129844 ↗
  • InjuryOct 20, 2023Material Rupture; Improper or Incorrect Procedure or Method; Material Rupture; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE PULMONIC VALVE VIA THE RIGHT JUGULAR VEIN ACCESS, DURING THE FIRST INFLATION PRIOR TO REACHING THE RATED BURST PRESSURE, THE PTA BALLOON ALLEGEDLY RUPTURED BEFORE OR AT 3 ATM AT MOST, AND ALLEGEDLY HAD A SMALL PINPOINT LEAKAGE IN THE DISTAL SHOULDER OF THE BALLOON. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED PULMONARY ARTERY DISSECTION AND HAD A PARTIAL RUPTURE OF THE PULMONARY ARTERY THAT CAUSED PERICARDIAL INFUSION. FURTHERMORE, THE PROCEDURE WAS UNABLE TO BE COMPLETED DUE TO THE RISK OF SUBSEQUENT BALLOON INFLATIONS AND THE PROCEDURE WAS ABORTED FOR PATIENT'S SAFETY. REPORTEDLY, THE PATIENT WAS KEPT FOR FORTY-EIGHT HOURS IN THE INTENSIVE CARE UNIT FOR CLOSE MONITORING OF THE PROGRESSIVE PULMONARY ARTERY RUPTURE AND HAD ACTIVITY RESTRICTIONS AT HOME FOR ABOUT TWO TO FOUR WEEKS. HOWEVER, THE PATIENT WAS RECOVERED WELL.

    Patient problems: Pericardial Effusion; Vascular Dissection

    MAUDE report 17974223 ↗
  • MalfunctionSep 29, 2023Leak/Splash; Material Rupture; Improper or Incorrect Procedure or Method; Leak/Splash; Material Rupture; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING A BALLOON VALVULOPLASTY PROCEDURE FOR VALVULAR PULMONIC STENOSIS, THROUGH THE RIGHT JUGULAR VEIN, THE DEVICE ALLEGEDLY HAD A LEAK BETWEEN SHAFT AND BALLOON INTERFACE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17841275 ↗
  • MalfunctionFeb 8, 2021Retraction Problem; Unraveled Material; Detachment of Device or Device Component; Retraction Problem; Unraveled Material; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE WAS ALLEGEDLY DIFFICULT TO BE PULLED BACK. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11290979 ↗
  • MalfunctionJul 19, 2019Break; Retraction Problem; Material Rupture; Break; Retraction Problem; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL ATG120244 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A RETRACTION PROBLEM AND RUPTURE. THERE WAS ONE PATIENT INVOLVED WITH NO REPORTED PATIENT INJURY. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8808057 ↗
  • MalfunctionOct 18, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THE PTA BALLOON ALLEGEDLY DETACHED FROM THE CATHETER. REPORTEDLY, THE HEALTH CARE PROVIDER WAS ABLE TO SUCCESSFULLY RETRIEVE THE BALLOON SEGMENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6960172 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates