Skip to content

Ultraverse® 018 U8150530Recalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse® 018 PTA Balloon Dilatation Catheter 5 mm x 300 mm, 150 cm Catheter · Class 2

All models of Ultraverse 018 (39)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741065194), received between Jun 13, 2025 and Apr 13, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material1
  • Device Dislodged or Dislocated1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2026Inflation Problem; Device Contamination with Chemical or Other Material

    On (b)(6)2026, a 48-year-old male patient underwent a percutaneous transluminal angioplasty (PTA) procedure using the Ultraverse 018 balloon catheter. During the procedure, the balloon catheter allegedly unable to inflate and the particles came from inside of the balloon. The procedure was completed using another balloon. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24856074 ↗
  • MalfunctionJun 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE 018 PTA BALLOON CATHETER VIA COMMON FEMORAL CONTRALATERAL APPROACH. DURING THE PROCEDURE, THE MARKER ON THE CATHETER BALLOON WAS ALLEGEDLY OFF A CENTIMETER OR TWO. IT WAS FURTHER REPORTED THAT THE BALLOON INFLATED BEYOND THE MARKER. BALLOON WAS REMOVED, AND ANGIOGRAM WAS PERFORMED TO VERIFY NO INJURY TO THE VESSEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22201310 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page