Skip to content

Ultraverse® 035 PTA Dilatation Catheter U3513066Recalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 6.0mm x 60mm balloon, 130cm shaft · Class 2

All models of Ultraverse 035 PTA Dilatation Catheter (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741093180), received between Jan 3, 2017 and Oct 14, 2024 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break2
  • Detachment of Device or Device Component2
  • Device-Device Incompatibility2
  • Failure to Advance2
  • Material Deformation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body In Patient1
  • No Patient Involvement1
  • Vascular Dissection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE RESULTS OF THE CLINICAL TRIAL THAT DURING PRE-DILATATION OF THE LEFT LEG NON-TARGET LESION STENOSIS ON THE PROXIMAL SUPERFICIAL FEMORAL ARTERY USING THE BALLOON CATHETER, THE SUBJECT HAD AN ADVERSE EVENT OF DISSECTION. REPORTEDLY, THE ADVERSE EVENT WAS NOT RELATED TO THE STUDY DEVICE BUT CAUSAL RELATIONSHIP TO THE STUDY PROCEDURE. REPORTEDLY, THE ADVERSE EVENT WAS TREATED WITH STENT PLACEMENT AND THE OUTCOME WAS RECOVERED. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Vascular Dissection

    MAUDE report 20437000 ↗
  • InjuryJan 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT POST AN ANGIOPLASTY PROCEDURE, THE PATIENT ALLEGEDLY HAD NO PULSATION AND LOW SKIN TEMPERATURE. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED MASSIVE THROMBOSIS IN THE ARTERIES OF THE LEFT LOWER EXTREMITY AND AN EMERGENCY ARTERIOTOMY AND EXPLORATORY THROMBECTOMY OF THE LEFT LOWER EXTREMITY ARTERIES WAS PERFORMED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18423707 ↗
  • InjuryMay 25, 2023Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation; Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CATHETER ALLEGEDLY HAD A KINK. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE ALLEGEDLY FAILED TO ADVANCE THROUGH THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17001145 ↗
  • MalfunctionOct 28, 2022Break; Failure to Advance; Device-Device Incompatibility; Material Deformation; Break; Failure to Advance; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE VIA FISTULA THROUGH RIGHT FEMORAL ARTERY, THE GUIDE WIRE ALLEGEDLY FAILED TO ADVANCE THROUGH THE LUMEN OF THE DEVICE CATHETER. IT WAS FURTHER REPORTED THAT THE DEVICE WAS ALLEGEDLY BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15690585 ↗
  • MalfunctionAug 18, 2021Break; Break

    IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE DEVICE WAS ALLEGEDLY BROKEN. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12334951 ↗
  • MalfunctionJan 3, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PREPARATION FOR AN ANGIOPLASTY PROCEDURE IN THE LEFT SFA THE BALLOON CATHETER ALLEGEDLY DETACHED FROM THE INFLATION HUB UPON REMOVING THE BALLOON FROM THE PROTECTIVE SLEEVE. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO REPORTED PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 6220109 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates