Skip to content

Ultraverse® 035 PTA Dilatation Catheter U35130610Recalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 6.0mm x 100mm balloon, 130cm shaft · Class 2

All models of Ultraverse 035 PTA Dilatation Catheter (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741093203), received between Oct 5, 2017 and Mar 24, 2025 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture7
  • Detachment of Device or Device Component3
  • Difficult to Remove2
  • Leak/Splash2
  • Material Split, Cut or Torn2
  • Retraction Problem2
  • Break1
  • Deflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2025Difficult to Remove; Stretched; Detachment of Device or Device Component; Difficult to Remove; Stretched; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY GOT STUCK INSIDE A SHEATH AFTER DEPLOYING AND FULLING DEFLATING THE BALLOON. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678169 ↗
  • MalfunctionJan 28, 2025Break; Material Rupture; Break; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, THE BALLOON ALLEGEDLY HAD A FLUID LEAK IN THE BODY DURING THE EXPANSION. IT WAS FURTHER REPORTED THAT THE BALLOON WAS TAKEN OUT OF THE BODY AND FOUND THAT THE BALLOON IS INDEED BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21251255 ↗
  • MalfunctionOct 18, 2024Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY DIFFICULT TO RETRACT AFTER DEFLATING. IT WAS FURTHER REPORTED THAT THE BALLOON ALLEGEDLY DETACHED. REPORTEDLY, THE BALLOON WAS PARTIALLY ATTACHED WITH GUIDEWIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20479434 ↗
  • MalfunctionJul 10, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, UPON FLUSHING THE BALLOON CATHETER, SALINE WAS ALLEGEDLY FOUND TO BE LEAKED FROM MID CATHETER SEGMENT ITSELF. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19708000 ↗
  • MalfunctionDec 9, 2021Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE SHAFT ON THE PTA BALLOON ALLEGEDLY LEAKED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12966369 ↗
  • MalfunctionJun 23, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED AT 8 ATM. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12048275 ↗
  • MalfunctionMar 23, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT LEG CLAUDICATION, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11548739 ↗
  • MalfunctionMar 22, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT LEG CLAUDICATION, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11537737 ↗
  • MalfunctionJan 15, 2020Deflation Problem; Retraction Problem; Device-Device Incompatibility; Deflation Problem; Retraction Problem; Device-Device Incompatibility

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U35130610 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED DEFLATION ISSUE AND DEVICE-DEVICE INCOMPATIBILITY. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. OF THE ONE DEFLATION ISSUE AND DEVICE-DEVICE INCOMPATIBILITY, ONE INVOLVED THE PATIENT WITH NO PATIENT CONSEQUENCES. THE PATIENT WAS A (B)(6) YEAR OLD MALE AND WEIGHED (B)(6) LBS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9593815 ↗
  • MalfunctionJan 14, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U35130610 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED PATIENT INJURY. THE MALE PATIENT'S AGE AND WEIGHT WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9588301 ↗
  • MalfunctionAug 22, 2018Retraction Problem; Material Rupture; Retraction Problem; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA, USING A CONTRALATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED AFTER THE SECOND INFLATION. IT WAS FURTHER REPORTED THAT THERE WAS DIFFICULTY RETRACTING THE BALLOON THROUGH THE SHEATH. THE BALLOON AND SHEATH WERE REMOVED TOGETHER AS ONE UNIT. THE PROCEDURE WAS COMPLETED WITH ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7806384 ↗
  • InjuryOct 5, 2017Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED BETWEEN 16 - 18 ATM ON THE SECOND INFLATION, SUBSEQUENTLY THE DISTAL PORTION OF THE BALLOON ALLEGEDLY DETACHED FROM THE CATHETER SHAFT. REPORTEDLY, THE HEALTH CARE PROVIDER DEPLOYED A COVERED STENT TO PIN THE DETACHED BALLOON SEGMENT TO THE VESSEL WALL. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6916340 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates