Ultraverse® 035 PTA Dilatation Catheter U35130630Recalls & reported problems
Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 6.0mm x 300mm balloon, 130cm shaft · Class 2
All models of Ultraverse 035 PTA Dilatation Catheter (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741093258), received between Nov 10, 2016 and Jun 13, 2023 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Retraction Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Retraction Problem4
- Break2
- Material Rupture2
- Deflation Problem1
- Difficult to Remove1
- Inflation Problem1
- Leak/Splash1
- Material Twisted/Bent1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 13, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY GOT STUCK WITHIN THE SHEATH WHEN DEFLATED TO PULL OUT OF THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17115949 ↗MalfunctionApr 13, 2023Break; Inflation Problem; Material Twisted/Bent; Break; Inflation Problem; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT SUPERFICIAL FEMORAL ARTERY THROUGH CONTRALATERAL APPROACH, THE PTA BALLOON WAS SEEMED TO BE ALLEGEDLY TWISTED WHILE IN PATIENT. IT WAS FURTHER REPORTED THAT WHILE INFLATING TO NOMINAL PRESSURE, THERE WAS ONE CENTIMETER SEGMENT WHICH WAS NOT DILATED. REPORTEDLY, THE BALLOON WAS RE POSITIONED TO POST DILATE THE SEGMENT THAT WAS NOTICED AS NOT DILATED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16730482 ↗MalfunctionOct 26, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U35130630 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE (B)(6) YEAR OLD MALE PATIENT WAS (B)(6) LBS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10737883 ↗MalfunctionJul 22, 2020Retraction Problem; Retraction Problem
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL U35130630 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED RETRACTION PROBLEM. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED WITH NO REPORTED PATIENT INJURY. THE FEMALE PATIENT WAS (B)(6) AND WEIGHT NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10309817 ↗MalfunctionJun 19, 2018Retraction Problem; Material Rupture; Retraction Problem; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA USING AN ANTEGRADE APPROACH, THE PTA BALLOON ALLEGEDLY GOT STUCK INSIDE THE 6FR SHEATH AND SUBSEQUENTLY COULD NOT BE RETRACTED. THE BALLOON AND SHEATH WERE REMOVED AS ONE UNIT. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7617463 ↗MalfunctionMay 29, 2018Deflation Problem; Retraction Problem; Deflation Problem; Retraction Problem
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THE PTA BALLOON ALLEGEDLY WAS UNABLE TO DEFLATE AND SUBSEQUENTLY WAS DIFFICULT TO REMOVE THROUGH THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7551429 ↗MalfunctionNov 10, 2016Break; Leak/Splash; Retraction Problem; Break; Leak/Splash; Retraction Problem
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA, A LEAK WAS ALLEGEDLY IDENTIFIED NEAR THE HUB. REPORTEDLY, DIFFICULTY WAS ENCOUNTERED DURING RETRACTION OF THE DEVICE; THEREFORE, A SYRINGE WAS USED TO REMOVE THE CONTRAST AND SALINE FROM THE BALLOON TO AIDE IN THE RETRACTION FROM THE SHEATH. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6093705 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page